TOWARD CONSENSUS ON QUANTITATIVE ASSESSMENT OF MEDICAL IMAGING-SYSTEMS

被引:98
作者
METZ, CE [1 ]
WAGNER, RF [1 ]
DOI, K [1 ]
BROWN, DG [1 ]
NISHIKAWA, RM [1 ]
MYERS, KJ [1 ]
机构
[1] US FDA,CTR DEVICES & RADIOL HLTH,OFF SCI & TECHNOL,ROCKVILLE,MD 20857
关键词
DETECTIVE QUANTUM EFFICIENCY (DQE); IDEAL OBSERVER; IMAGE QUALITY; NOISE EQUIVALENT QUANTA (NEQ); RECEIVER OPERATING CHARACTERISTIC (ROC) ANALYSIS;
D O I
10.1118/1.597511
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Consensus has been developing over the past few decades on a number of measurements required for the laboratory assessment of medical imaging modalities. Nevertheless, understanding of the connection between these measurements and human observer performance in a broad range of tasks remains far from complete. Focusing primarily on projection radiography to provide concrete examples, this overview indicates areas in which consensus on methodology for physical image-quality measurement has been established. Concepts such as ''noise equivalent quanta'' (NEQ) and ''detective quantum efficiency'' (DQE) have been found useful for normalizing physical measurements on an absolute scale and for relating those measurements to the decision performance of a hypothetical ''ideal observer'' that effectively performs decision tasks from the image data. The connection between ideal observer performance and human performance, as determined by receiver operating characteristic (ROC) analysis, remains to be understood for many clinically relevant tasks.
引用
收藏
页码:1057 / 1061
页数:5
相关论文
共 47 条