LIQUID-CRYSTAL SKIN THERMOMETRY - AN ACCURATE REFLECTION OF CORE TEMPERATURE

被引:9
作者
BRULL, SJ [1 ]
CUNNINGHAM, AJ [1 ]
CONNELLY, NR [1 ]
OCONNOR, TZ [1 ]
SILVERMAN, DG [1 ]
机构
[1] YALE UNIV,NEW HAVEN HOSP,SCH MED,DEPT ANESTHESIOL,NEW HAVEN,CT 06510
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1993年 / 40卷 / 04期
关键词
MEASUREMENT TECHNIQUES; LIQUID CRYSTAL THERMOMETRY; SURGERY; CARDIOPULMONARY BYPASS; TEMPERATURE; MONITORING;
D O I
10.1007/BF03009638
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Oesophageal, rectal, bladder, tympanic and pulmonary artery sites are used intraoperatively to measure body temperature. However, the temperatures measured at each site have different physiological and practical importance. The present two-part study sought to compare liquid crystal (CR) skin temperature with other temperature monitors which are used routinely during surgery. The first part compared CR with oesophageal (OS) temperature during general inhalational anaesthesia. The second part compared CR with OS, pulmonary artery (PA), and bladder (BL) temperatures during the periods of rapid temperature change associated with cardiopulmonary bypass (CPB). During the first part, the mean difference between OS and CR was 0.14+/-0.85-degrees-C; this difference remained consistent over time (P < 0.05 by repeated measures analysis of variance). During the second part, the difference in temperature readings between CR and each of the other monitors remained consistent except for CR vs PA and CR vs OS during the cooling period of CPB, when the iced cardioplegia slush directly affected the PA and OS temperatures. This study suggests that CR, an inexpensive and noninvasive means of temperature monitoring, reflects trends in temperature changes in the clinical setting.
引用
收藏
页码:375 / 381
页数:7
相关论文
共 13 条
[1]   DOES FOREHEAD LIQUID-CRYSTAL TEMPERATURE ACCURATELY REFLECT CORE TEMPERATURE [J].
ALLEN, GC ;
HORROW, JC ;
ROSENBERG, H .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1990, 37 (06) :659-662
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   CONTINUOUS MONITORING OF SKIN TEMPERATURE USING A LIQUID-CRYSTAL THERMOMETER DURING ANESTHESIA [J].
BURGESS, GE ;
COOPER, JR ;
MARINO, RJ ;
PEULER, MJ .
SOUTHERN MEDICAL JOURNAL, 1978, 71 (05) :516-518
[4]  
CORK RC, 1983, ANESTH ANALG, V62, P211
[5]   BODY-TEMPERATURE AND ANESTHESIA [J].
HALL, GM .
BRITISH JOURNAL OF ANAESTHESIA, 1978, 50 (01) :39-44
[6]   LIQUID-CRYSTAL THERMOMETRY DURING ANESTHESIA [J].
LACOUMENTA, S ;
HALL, GM .
ANAESTHESIA, 1984, 39 (01) :54-56
[7]  
LEES DE, 1978, ANESTH ANALG, V57, P669
[8]   MORPHINE AND POSTOPERATIVE REWARMING IN CRITICALLY ILL PATIENTS [J].
RODRIGUEZ, JL ;
WEISSMAN, C ;
DAMASK, MC ;
ASKANAZI, J ;
HYMAN, AI ;
KINNEY, JM .
CIRCULATION, 1983, 68 (06) :1238-1246
[9]  
ROSENBERG H, 1992, CLIN ANESTH, P589
[10]   THE THERMOREGULATORY THRESHOLD IN HUMANS DURING HALOTHANE ANESTHESIA [J].
SESSLER, DI ;
OLOFSSON, CI ;
RUBINSTEIN, EH ;
BEEBE, JJ .
ANESTHESIOLOGY, 1988, 68 (06) :836-842