PERFORMANCE OF A NEUTRON POLARIMETER TO MEASURE THE ELECTRIC FORM-FACTOR OF THE NEUTRON

被引:11
作者
EDEN, T
MADEY, R
ZHANG, WM
ANDERSON, BD
BALDWIN, AR
CAMERON, JM
FINN, JM
FOSTER, CC
KELLY, JJ
KOWALSKI, S
LOURIE, R
MARKOWITZ, P
NI, B
PELLA, PJ
SCHAMBACH, J
SPRAKER, M
STEINFELDS, E
STEPHENSON, EJ
机构
[1] HAMPTON UNIV, HAMPTON, VA 23668 USA
[2] INDIANA UNIV, CYCLOTRON FACIL, BLOOMINGTON, IN 47405 USA
[3] COLL WILLIAM & MARY, WILLIAMSBURG, VA 23185 USA
[4] UNIV MARYLAND, COLL PK, MD 20742 USA
[5] MIT, CAMBRIDGE, MA 02139 USA
[6] UNIV VIRGINIA, CHARLOTTESVILLE, VA 22904 USA
[7] GETTYSBURG COLL, GETTYSBURG, PA 17325 USA
基金
美国国家科学基金会;
关键词
D O I
10.1016/0168-9002(94)91327-7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We measured the analyzing power and the efficiency of a neutron polarimeter that was designed and constructed to measure the neutron electric form factor G(E)(n). The measurement was carried out at the Indiana University Cyclotron Facility (IUCF) with the C-14(($) over right arrow p, ($) over right arrow n)N-14 reaction at a proton beam energy of 138.4 MeV. The analyzing power, (A(y)), averaged over the acceptance of the polarimeter is typically 0.44+/-0.02(-0.33)(+0.04) at a neutron energy of 135 MeV. With a lead-steel wall, consisting of 4 in. of lead sandwiched between two 1 3/8-in. steel plates, ahead of the polarimeter, the average analyzing power is the same within statistics, namely, (A(y)) = 0.43+/-0.02(-0.03)(+0.04). The large systematic uncertainty quoted here is dominated by an uncertainty from the inability at the time of the experiment of measuring the longitudinal polarization component of the sideways polarized protons; this systematic uncertainty can be reduced significantly to 0.02 in any future calibration at the IUCF. In each case, the efficiency of the polarimeter is about 0.26% for 135 MeV neutrons at a pulse-height threshold of 4 MeV equivalent electron energy (MeVee) on the front detectors and 10 MeVee on the rear detectors. This analysis reveals a small analyzing power from the carbon in the front-analyzer detectors.
引用
收藏
页码:432 / 441
页数:10
相关论文
共 15 条
[1]   COMPARISON OF O-DEGREES (P,N) CROSS-SECTIONS AND B (M1) VALUES FOR SEPARATING CURRENT AND SPIN CONTRIBUTIONS TO ISOVECTOR M1 TRANSITIONS [J].
ANDERSON, BD ;
MCCARTHY, RJ ;
AHMAD, M ;
FAZELY, A ;
KALENDA, AM ;
KNUDSON, JN ;
WATSON, JW ;
MADEY, R ;
FOSTER, CC .
PHYSICAL REVIEW C, 1982, 26 (01) :8-13
[2]   NUCLEON-NUCLEON-SCATTERING ANALYSES .2. NEUTRON-PROTON SCATTERING FROM 0 TO 425 MEV AND PROTON-PROTON SCATTERING FROM 1 TO 500 MEV [J].
ARNDT, RA ;
HACKMAN, RH ;
ROPER, LD .
PHYSICAL REVIEW C, 1977, 15 (03) :1002-1020
[3]  
ARNDT RA, UNPUB SCATTERING ANA
[4]   POLARIZATION TRANSFER IN ELASTIC ELECTRON-SCATTERING FROM NUCLEONS AND DEUTERONS [J].
ARNOLD, RG ;
CARLSON, CE ;
GROSS, F .
PHYSICAL REVIEW C, 1981, 23 (01) :363-374
[5]  
BARSCHALL HH, 1970, POLARIZATION PHENOME
[6]   MINUIT - SYSTEM FOR FUNCTION MINIMIZATION AND ANALYSIS OF PARAMETER ERRORS AND CORRELATIONS [J].
JAMES, F ;
ROOS, M .
COMPUTER PHYSICS COMMUNICATIONS, 1975, 10 (06) :343-367
[7]  
LOW DA, 1983, INDIANA U CYCLOTRON, P156
[8]   POLARIMETERS FOR MEDIUM-ENERGY NEUTRONS [J].
MADEY, R ;
BALDWIN, AR ;
PELLA, PJ ;
SCHAMBACH, J ;
SELLERS, RM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :231-240
[9]   LARGE VOLUME NEUTRON DETECTORS WITH SUBNANOSECOND TIME DISPERSIONS [J].
MADEY, R ;
WATSON, JW ;
AHMAD, M ;
ANDERSON, BD ;
BALDWIN, AR ;
CASSON, AL ;
CASSON, W ;
CECIL, RA ;
FAZELY, A ;
KNUDSON, JM ;
LEBO, C ;
PAIRSUWAN, W ;
PELLA, PJ ;
VARGA, JC ;
WITTEN, TR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 214 (2-3) :401-413
[10]  
MADEY R, 1987 1988 BAT LIN AC, P140