FLOW SUPPORT CATHETER FOR PROLONGED MAINTENANCE OF CORONARY BLOOD-FLOW

被引:4
作者
CLUGSTON, R
OESTERLE, SN
MATTHEWS, R
DAWSON, R
GARRISON, M
ALKER, K
REEVES, T
VIRMANI, R
KLONER, RA
机构
[1] HOSP GOOD SAMARITAN,INST HEART,LOS ANGELES,CA 90017
[2] UNIV CALIF LOS ANGELES,LOS ANGELES CTY HARBOR MED CTR,TORRANCE,CA 90509
[3] ARMED FORCES INST PATHOL,WASHINGTON,DC 20306
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1991年 / 24卷 / 04期
关键词
CORONARY ANGIOPLASTY; INTRACORONARY STENT; REGIONAL MYOCARDIAL BLOOD FLOW;
D O I
10.1002/ccd.1810240419
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A newly designed flow support catheter with a supporting wire mesh cage which can be expanded into a tubular configuration and then readily reduced was evaluated in mongrel dogs. Regional myocardial blood flow (RMBF) was measured using the radioactive microsphere technique in the area of both balloon-denuded instrumented and control non-instrumented coronary arteries following placement of either a fixed-wire or a higher profile rapid exchange flow support catheter. At 5, 20, and 180 min following delivery and expansion of either device, RMBF was not significantly different in left ventricular subepicardium and subendocardium perfused by the instrumented vs. the control coronary arteries. Angiography demonstrated widely patent instrumented arteries in 15/18 dogs; in no dog was side branch occlusion observed. Significant cage thrombus deposition was seen angiographically in 3 animals causing temporary total coronary occlusion in 1. Following reduction and removal of the flow support catheter, vessel patency was present in all dogs. The flow support catheter is an effective endovascular stenting device capable of providing structural arterial support, while simultaneously maintaining distal coronary blood flow. It is envisioned that the primary application of this catheter will be to enable primary salvage of vessels acutely injured during coronary angioplasty, by "tacking up" intimal flaps for an extended period. It may also provide a bridge to emergency surgical revascularization.
引用
收藏
页码:308 / 314
页数:7
相关论文
共 16 条
[1]   EFFECT OF VERAPAMIL ON INFARCT SIZE IN DOGS SUBJECTED TO CORONARY-ARTERY OCCLUSION WITH TRANSIENT REPERFUSION [J].
CAMPBELL, CA ;
KLONER, RA ;
ALKER, KJ ;
BRAUNWALD, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 8 (05) :1169-1174
[2]   CATHETER REPERFUSION FOR FAILED EMERGENCY CORONARY ANGIOPLASTY WITHOUT SUBSEQUENT BYPASS-SURGERY [J].
CIAMPRICOTTI, R ;
DEKKERS, PJWM ;
ELGAMAL, MIH ;
VANDERKRIEKEN, AM ;
RELIK, THFM .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1989, 18 (03) :159-164
[3]   EMERGENCY CORONARY-BYPASS SURGERY AFTER CORONARY ANGIOPLASTY - THE NATIONAL-HEART-LUNG-AND-BLOOD-INSTITUTES PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY REGISTRY EXPERIENCE [J].
COWLEY, MJ ;
DORROS, G ;
KELSEY, SF ;
VANRADEN, M ;
DETRE, KM .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (12) :C22-C26
[4]   ANGIOGRAPHIC AND CLINICAL PREDICTORS OF ACUTE CLOSURE AFTER NATIVE VESSEL CORONARY ANGIOPLASTY [J].
ELLIS, SG ;
ROUBIN, GS ;
KING, SB ;
DOUGLAS, JS ;
WEINTRAUB, WS ;
THOMAS, RG ;
COX, WR .
CIRCULATION, 1988, 77 (02) :372-379
[5]   INTRACORONARY STENTS - WILL THEY FULFILL THEIR PROMISE AS AN ADJUNCT TO ANGIOPLASTY [J].
ELLIS, SG ;
TOPOL, EJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (06) :1425-1430
[6]  
GASPARD P E, 1990, Journal of the American College of Cardiology, V15, p118A
[7]   TRANS-LUMINAL INTRACORONARY REPERFUSION CATHETER - A DEVICE TO MAINTAIN CORONARY PERFUSION BETWEEN FAILED CORONARY ANGIOPLASTY AND EMERGENCY CORONARY-BYPASS SURGERY [J].
HINOHARA, T ;
SIMPSON, JB ;
PHILLIPS, HR ;
STACK, RS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (05) :977-982
[8]  
HODES Z I, 1989, Journal of the American College of Cardiology, V13, p155A
[9]   REMOVAL OF AN OCCLUSIVE CORONARY DISSECTION FLAP BY USING AN ATHERECTOMY CATHETER [J].
LEE, TC ;
HARTZLER, GO ;
RUTHERFORD, BD ;
MCCONAHAY, DR .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 20 (03) :185-188
[10]   CLINICAL AND ANGIOGRAPHIC RESULTS OF BALLOON-EXPANDABLE INTRACORONARY STENTS IN RIGHT CORONARY-ARTERY STENOSES [J].
LEVINE, MJ ;
LEONARD, BM ;
BURKE, JA ;
NASH, ID ;
SAFIAN, RD ;
DIVER, DJ ;
BAIM, DS .