PHOTOGRAMMETRIC ACCURACY MEASUREMENTS OF HEAD HOLDER SYSTEMS USED FOR FRACTIONATED RADIOTHERAPY

被引:76
作者
MENKE, M
HIRSCHFELD, F
MACK, T
PASTYR, O
STURM, V
SCHLEGEL, W
机构
[1] BODENSEEWERK GERATETECH GMBH,D-88641 UBERLINGEN,GERMANY
[2] UNIV KLIN KOLN,STEREOTAXIE & FUNKT NEUROCHIRURG ABT,D-50931 COLOGNE,GERMANY
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1994年 / 29卷 / 05期
关键词
FRACTIONATED RADIOTHERAPY; PATIENT MOVEMENT; PATIENT POSITIONING; PHOTOGRAMMETRY;
D O I
10.1016/0360-3016(94)90412-X
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: We describe how stereo photogrammetry can be used to determine immobilization and repositioning accuracies of head holder systems used for fractionated radiotherapy of intracranial lesions. Methods and Materials: The apparatus consists of two video cameras controlled by a personal computer and a bite block based landmark system. Position and spatial orientation of the landmarks are monitored by the cameras and processed for the real-time calculation of a target point's actual position relative to its initializing position. The target's position is assumed to be invariant with respect to the landmark system. We performed two series of 30 correlated head motion measurements on two test persons. One of the series was done with a thermoplastic device, the other one with a cast device developed for stereotactic treatment at the German Cancer Research Center. Immobilization and repositioning accuracies were determined with respect to a target point situated near the base of the skull. The repositioning accuracies were described in terms of the distributions of the mean displacements of the single motion measurements. Results: Movements of the target in the order of 0.05 mm caused by breathing could be detected with a maximum resolution in time of 12 ms. The data derived from the investigation of the two test persons indicated similar immobilization accuracies for the two devices, but the repositioning errors were larger for the thermoplastic device than for the cast device. Apart from this, we found that for the thermoplastic mask the lateral repositioning error depended on the order in which the mask was closed. Conclusion: The photogrammetric apparatus is a versatile tool for accuracy measurements of head holder devices used for fractionated radiotherapy.
引用
收藏
页码:1147 / 1155
页数:9
相关论文
共 11 条
  • [1] GALL KP, 1993, MED PHYS, V20, P1153, DOI 10.1118/1.596969
  • [2] MEASUREMENT OF PATIENT MOVEMENT DURING RADIATION-THERAPY
    KELSEY, CA
    CONNOR, WG
    LANE, RG
    [J]. RADIOLOGY, 1972, 103 (03) : 697 - &
  • [3] KRAUS K, 1990, GRUNDLAGEN STANDARD, V1
  • [4] PLASTIC MATERIAL USED TO OPTIMIZE RADIOTHERAPY OF HEAD AND NECK TUMORS AND THE MAMMARY-CARCINOMA
    NIEWALD, M
    LEHMANN, W
    UHLMANN, U
    SCHNABEL, K
    LEETZ, HK
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 11 (01) : 55 - 63
  • [5] NIEWALD M, 1986, STRAHLENTHER ONKOL, V162, P448
  • [6] PATIENT MOVEMENTS DURING RADIOTHERAPY
    NORWOOD, HM
    STUBBS, B
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (674) : 155 - 158
  • [7] STEREOTACTICALLY GUIDED CONVERGENT BEAM IRRADIATION WITH A LINEAR-ACCELERATOR - LOCALIZATION-TECHNIQUE
    PASTYR, O
    HARTMANN, GH
    SCHLEGEL, W
    SCHABBERT, S
    TREUER, H
    LORENZ, WJ
    STURM, V
    [J]. ACTA NEUROCHIRURGICA, 1989, 99 (1-2) : 61 - 64
  • [8] COMPUTER-SYSTEMS AND MECHANICAL TOOLS FOR STEREOTAXICALLY GUIDED CONFORMATION THERAPY WITH LINEAR ACCELERATORS
    SCHLEGEL, W
    PASTYR, O
    BORTFELD, T
    BECKER, G
    SCHAD, L
    GADEMANN, G
    LORENZ, WJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (04): : 781 - 787
  • [9] STOER J, 1983, EINFUHRUNG NUMERISCH, V1, P224
  • [10] STEREOTACTIC COMPUTER-TOMOGRAPHY WITH A MODIFIED RIECHERT-MUNDINGER DEVICE AS THE BASIS FOR INTEGRATED STEREOTACTIC NEURORADIOLOGICAL INVESTIGATIONS
    STURM, V
    PASTYR, O
    SCHLEGEL, W
    SCHARFENBERG, H
    ZABEL, HJ
    NETZEBAND, G
    SCHABBERT, S
    BERBERICH, W
    [J]. ACTA NEUROCHIRURGICA, 1983, 68 (1-2) : 11 - 17