HIGH-SENSITIVITY GAS-PHASE SEQUENCE-ANALYSIS OF PROTEINS AND PEPTIDES ON PVDF MEMBRANES USING SHORT CYCLE TIMES

被引:9
作者
REIM, DF [1 ]
SPEICHER, DW [1 ]
机构
[1] WISTAR INST ANAT & BIOL,ROOM C102,3601 SPRUCE ST,PHILADELPHIA,PA 19104
关键词
D O I
10.1006/abio.1993.1460
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An optimized sequencer program with a cycle time of 38 min which is specifically tailored for analysis using polyvinylidene difluoride (PVDF) membranes has been developed. The program was developed using a pulsed liquid-phase instrument which was converted to gas-phase acid delivery. Gas-phase acid delivery minimized sample extraction from PVDF membranes and improved tryptophan yields in at least some cases. Other modifications which contributed to reliable high sensitivity sequencer performance included use of a Blott cartridge, substitution of ethyl acetate:heptane (1:1, v/v) instead of butyl chloride as the extraction solvent, use of a modified 100-μl injection loop with an internal restrictor to reliably inject nearly 90% of the sample, and an HPLC gradient which resolved tryptophan from diphenylurea. These shortened cycle times were achieved at the conventional gas-phase reaction temperature. A slight increase in lag or carryover at prolines was compensated by reduced background from nonspecific acid cleavage which facilitated extended and/or high sensitivity sequencing of large proteins. Reproducible high initial and repetitive cycle yields were obtained with a wide range of experimental peptides which were electroblotted from either 1D or 2D polyacrylamide gels onto high retention PVDF-membranes. Initial yields of the majority of the experimental samples analyzed with this program were less than 5 pmol. In addition, most samples with initial yields below 1-2 pmol yielded sufficient sequence information to identify the protein by comparison to protein sequence databases or to design oligonucleotide probes. © 1994 Academic Press, Inc. All rights reserved.
引用
收藏
页码:87 / 95
页数:9
相关论文
共 19 条
[1]   INTERNAL AMINO-ACID SEQUENCE-ANALYSIS OF PROTEINS SEPARATED BY ONE-DIMENSIONAL OR TWO-DIMENSIONAL GEL-ELECTROPHORESIS AFTER INSITU PROTEASE DIGESTION ON NITROCELLULOSE [J].
AEBERSOLD, RH ;
LEAVITT, J ;
SAAVEDRA, RA ;
HOOD, LE ;
KENT, SBH .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1987, 84 (20) :6970-6974
[2]  
CRIMMINS DL, 1992, TECHNIQUES PROTEIN C, V3, P35
[3]  
ECKERSKORN C, 1990, ELECTROPHORESIS, V11, P581
[4]   A PROTEIN SEQUENATOR [J].
EDMAN, P ;
BEGG, G .
EUROPEAN JOURNAL OF BIOCHEMISTRY, 1967, 1 (01) :80-&
[5]   INTERNAL PROTEIN-SEQUENCE ANALYSIS - ENZYMATIC DIGESTION FOR LESS THAN 10 MU-G OF PROTEIN-BOUND TO POLYVINYLIDENE DIFLUORIDE OR NITROCELLULOSE MEMBRANES [J].
FERNANDEZ, J ;
DEMOTT, M ;
ATHERTON, D ;
MISCHE, SM .
ANALYTICAL BIOCHEMISTRY, 1992, 201 (02) :255-264
[6]  
LEGENDRE N, 1990, Biotechniques, V9, P788
[7]  
MATSUDAIRA P, 1987, J BIOL CHEM, V262, P10035
[8]   MICROSEQUENCE ANALYSIS OF ELECTROBLOTTED PROTEINS .1. COMPARISON OF ELECTROBLOTTING RECOVERIES USING DIFFERENT TYPES OF PVDF MEMBRANES [J].
MOZDZANOWSKI, J ;
SPEICHER, DW .
ANALYTICAL BIOCHEMISTRY, 1992, 207 (01) :11-18
[9]   A METHOD FOR THE UNAMBIGUOUS IDENTIFICATION OF TRYPTOPHAN IN AUTOMATED PROTEIN-SEQUENCE ANALYSIS [J].
MULLERMICHEL, T ;
BOHLEN, P .
ANALYTICAL BIOCHEMISTRY, 1990, 191 (01) :169-173
[10]  
NIKA H, 1992, NEW APPROACHES SEQUE, P29