DOUBLE-BEAM BENT LAUE MONOCHROMATOR FOR CORONARY ANGIOGRAPHY

被引:21
作者
ILLING, G
HEUER, J
REIME, B
LOHMANN, M
MENK, RH
SCHILDWACHTER, L
DIX, WR
GRAEFF, W
机构
[1] Hamburger Synchrotronstrahlungslabor HASYLAB, Deutsches Elektronensynchrotron DESY, 22603 Hamburg
关键词
D O I
10.1063/1.1145981
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
High photon fluxes are crucial in dichromatic digital subtraction angiography with line scan systems to allow for adequate image quality. To obtain images wide enough and with sufficient vertical resolution, each of the two beams used has to be at least 100 mm wide but not more than 0.5 mm high at the patient's heart with the beam energies bracketing the iodine-K edge at 33.17 keV. A new monochromator in Laue transmission geometry has been developed, successfully tested at beamline ID11 of the European Synchrotron Radiation Facility (ESRF) and is now in use at the angiography wiggler beamline W2 at HASYLAB. The monochromator consists of two almost identical goniometer heads which are independently controlled. Within each head a bent silicon single crystal extracts a quasimonochromatic beam and focuses it vertically to the patient thus also increasing intensity compared to an unbent crystal. With a storage ring current of 50 mA the new monochromator provides 0.8×10 11 photons/mm2/s in front of the patient which is an increase by a factor of 6 compared to the formerly used Bragg design with Ge crystals. Also, it is remarkably more stable against variations of the input power load. © 1995 American Institute of Physics.
引用
收藏
页码:1379 / 1381
页数:3
相关论文
共 6 条
[1]   CORONARY ANGIOGRAPHY USING SYNCHROTRON RADIATION - STUDIES IN HUMAN-SUBJECTS WITH THE SYSTEM NIKOS-II [J].
DIX, WR ;
ENGELKE, K ;
GRAEFF, W ;
HAMM, C ;
HEUER, J ;
KAEMPF, B ;
KUPPER, W ;
LOHMANN, M ;
REIME, B ;
REUMANN, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 314 (02) :307-315
[2]  
DIX WR, 1994, IN PRESS FRONTIERS S
[3]   X-RAY REFLECTIVITY OF BENT PERFECT CRYSTALS IN BRAGG AND LAUE GEOMETRY [J].
EROLA, E ;
ETELANIEMI, V ;
SUORTTI, P ;
PATTISON, P ;
THOMLINSON, W .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1990, 23 :35-42
[4]  
MENK RH, 1995, REV SCI INSTRUM 2, V66
[5]  
SUORTTI P, 1989, NUCL INSTR METH PH A, V269, P639
[6]  
SUORTTI P, 1993, NUCL INSTRUM METH A, V336, P301