HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF OFLOXAXIN IN SERUM

被引:14
作者
OHKUBO, T
KUDO, M
SUGAWARA, K
机构
[1] Department of Pharmacy, Hirosaki University Hospital
关键词
HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; OFLOXACIN; COLUMN-SWITCHING TECHNIQUE; PHARMACOKINETICS;
D O I
10.2116/analsci.7.741
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A high-performance liquid chromatographic method was developed for the assay of ofloxacin in serum. The serum samples were filtered through a Molcut II membrane filter after partial deproteinization in order to remove proteins. The filtrate was preseparated on a reversed-phase precolumn and introduced onto a reversed-phase analytical column by applying column-switching techniques. Ofloxacin and nalidixic acid as internal standards were detected by absorbance measurements at 300 nm. Using this system the concentration of ofloxacin could be determined at a limit of the order of 20 ng/ml, with a linear dynamic range of 80 - 4000 ng/ml. This procedure was applied to pharmacokinetic studies of healthy volunteers treated with ofloxacin.
引用
收藏
页码:741 / 743
页数:3
相关论文
共 7 条
[1]   SIMULTANEOUS DETERMINATION OF OFLOXACIN, FENBUFEN AND FELBINAC IN RAT PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KATAGIRI, Y ;
NAORA, K ;
ICHIKAWA, N ;
HAYASHIBARA, M ;
IWAMOTO, K .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 431 (01) :135-142
[2]   DETERMINATION OF OFLOXACIN IN PLASMA AND URINE BY LIQUID-CHROMATOGRAPHY [J].
LECOGUIC, A ;
BIDAULT, R ;
FARINOTTI, R ;
DAUPHIN, A .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 434 (01) :320-323
[3]   THE PHARMACOKINETICS AND TISSUE PENETRATION OF OFLOXACIN [J].
LOCKLEY, MR ;
WISE, R ;
DENT, J .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 1984, 14 (06) :647-652
[4]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF OFLOXACIN IN PLASMA AND URINE [J].
MIGNOT, A ;
LEFEBVRE, MA ;
FOURTILLAN, JB .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 430 (01) :192-197
[5]  
OCHIAI K, 1988, JPN J HOSP PHARM, V14, P323
[6]   DETERMINATION OF MITOMYCIN-C IN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH DUAL-ELECTRODE COULOMETRIC DETECTION [J].
OHKUBO, T ;
NAMBARA, T .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 527 (02) :441-446
[7]   INVITRO AND INVIVO ACTIVITY OF DL-8280, A NEW OXAZINE DERIVATIVE [J].
SATO, K ;
MATSUURA, Y ;
INOUE, M ;
UNE, T ;
OSADA, Y ;
OGAWA, H ;
MITSUHASHI, S .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1982, 22 (04) :548-553