A BILIRUBIN BIOSENSOR BASED ON A MULTILAYER NETWORK ENZYME ELECTRODE

被引:85
作者
SHOHAM, B [1 ]
MIGRON, Y [1 ]
RIKLIN, A [1 ]
WILLNER, I [1 ]
TARTAKOVSKY, B [1 ]
机构
[1] SAVYON DIAGNOST LTD,IL-77101 ASHDOD,ISRAEL
关键词
AMPEROMETRIC BIOSENSOR; BILIRUBIN; ENZYME MONOLAYER; ENZYME ELECTRODE; ELECTRON TRANSFER IN PROTEINS; ENZYME MULTILAYERS; BILIRUBIN OXIDASE; BIOSENSOR;
D O I
10.1016/0956-5663(95)96852-P
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper describes the construction of a bilirubin oxidase multilayer electrode that responds amperometrically to bilirubin concentration. The enzyme electrode was constructed by the covalent attachment of bilirubin oxidase layers to a self-assembled monolayer of 3-mercaptopropionate ester, which was associated with a gold surface. Electron transfer from the enzyme redox site to the electrode was mediated by ferrocene carboxylic acid. The multilayer electrode responded linearly to bilirubin concentration in solution. In contrast to the native enzyme, which is unstable, the bilirubin oxidase multilayer electrode was stable, and its shelf-life was more than three months at 4 degrees C. The amperometric response of the enzyme electrode towards bilirubin was dependent on the albumin concentration in the system. By adjustment of this concentration to the value present in newborns' serum, a calibration curve for the amperometric response as a function of bilirubin was deduced. The concentration of bilirubin in serum samples from newborns was determined.
引用
收藏
页码:341 / 352
页数:12
相关论文
共 25 条
[11]  
LOWRY OH, 1951, J BIOL CHEM, V193, P265
[12]  
Maisels M.J., 1989, NEONATOLOGY PATHOPHY
[13]   AMPEROMETRIC BIOSENSOR FOR RAPID MEASUREMENT OF 3-HYDROXYBUTYRATE IN UNDILUTED WHOLE-BLOOD AND PLASMA [J].
MCNEIL, CJ ;
SPOORS, JA ;
COOPER, JM ;
ALBERTI, KGMM ;
MULLEN, WH .
ANALYTICA CHIMICA ACTA, 1990, 237 (01) :99-105
[14]   KINETICS OF BILIRUBIN OXIDASE AND MODELING OF AN IMMOBILIZED BILIRUBIN OXIDASE REACTOR FOR BILIRUBIN DETOXIFICATION [J].
MULLON, CJP ;
KLIBANOV, AM ;
LANGER, R .
BIOTECHNOLOGY AND BIOENGINEERING, 1988, 31 (06) :536-546
[15]  
SCHELLER FW, 1989, BIOMED BIOCHIM ACTA, V48, P891
[16]   AN IMMOBILIZED ENZYME REACTOR FOR THE DETOXIFICATION OF BILIRUBIN [J].
SUNG, C ;
LAVIN, A ;
KLIBANOV, AM ;
LANGER, R .
BIOTECHNOLOGY AND BIOENGINEERING, 1986, 28 (10) :1531-1539
[17]  
Ulman A., 1991, INTRO ULTRATHIN ORGA
[18]   AN ULTRAMICROMETHOD FOR DETERMINATION OF CONJUGATED AND TOTAL BILIRUBIN IN SERUM OR PLASMA [J].
WALTERS, MI ;
GERARDE, HW .
MICROCHEMICAL JOURNAL, 1970, 15 (02) :231-&
[19]   A POLISHABLE AMPEROMETRIC BIOSENSOR FOR BILIRUBIN [J].
WANG, J ;
OZSOZ, M .
ELECTROANALYSIS, 1990, 2 (08) :647-650
[20]  
WENNBERG RP, 1979, CLIN CHEM, V25, P1444