POSITION-SENSITIVE SCINTILLATION DETECTOR FOR GAMMA-RAYS USING A GSO CRYSTAL

被引:10
作者
NAGAI, Y [1 ]
HYODO, T [1 ]
机构
[1] UNIV TOKYO,COLL ARTS & SCI,INST PHYS,MEGURO KU,TOKYO 153,JAPAN
关键词
D O I
10.1016/0168-9002(94)90635-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report the development of a gamma-ray imager using a 5 mm thick, 50 mm diameter Gd2SiO5 (GSO) crystal coupled to a 2D position-sensitive photomultiplier tube (Hamamatsu R2486). A considerable improvement in the position linearity and resolution was achieved upon placing around the GSO scintillator a glass ring which has the same thickness and the same index of refraction as the GSO crystal. A position resolution of 2.2 mm FWHM was obtained at 511 keV.
引用
收藏
页码:285 / 288
页数:4
相关论文
共 8 条
[1]   A POSITION-SENSITIVE PHOSWICH [J].
BIRD, AJ ;
ZHONG, H ;
RAMSDEN, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :332-336
[2]   AN INEXPENSIVE, HARD X-RAY-IMAGING SPECTROMETER FOR USE IN X-RAY ASTRONOMY AND ATOMIC PHYSICS [J].
HAILEY, CJ ;
HARRISON, F ;
LUPTON, JH ;
ZIOCK, KP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 276 (1-2) :340-346
[3]  
HE Z, 1992 IEEE NUCL SCI S, P129
[4]   POSITION-SENSITIVE PHOTOMULTIPLIER TUBES FOR SCINTILLATION IMAGING [J].
KUME, H ;
MURAMATSU, S ;
IIDA, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :359-363
[5]   POSITION-SENSITIVE SCINTILLATION DETECTOR FOR HARD X-RAYS [J].
POULSEN, JM ;
VERBENI, R ;
FRONTERA, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :398-402
[6]  
Saito H., 1992, Materials Science Forum, V105-110, P1967, DOI 10.4028/www.scientific.net/MSF.105-110.1967
[7]   CERIUM-ACTIVATED GD2SIO5 SINGLE-CRYSTAL SCINTILLATOR [J].
TAKAGI, K ;
FUKAZAWA, T .
APPLIED PHYSICS LETTERS, 1983, 42 (01) :43-45
[8]   DESIGN OF A MOSAIC BGO DETECTOR SYSTEM FOR POSITRON CT [J].
UCHIDA, H ;
YAMASHITA, T ;
IIDA, M ;
MURAMATSU, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :464-467