PULSED AXIAL ACTIVATION IN THE ION TRAP - A NEW METHOD FOR PERFORMING TANDEM MASS-SPECTROSCOPY (MS/MS)

被引:38
作者
LAMMERT, SA [1 ]
COOKS, RG [1 ]
机构
[1] PURDUE UNIV,DEPT CHEM,W LAFAYETTE,IN 47907
关键词
D O I
10.1002/rcm.1290060809
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A short DC voltage pulse is applied to one endcap of a quadrupole ion-trap mass spectrometer to activate trapped ions. High internal-energy deposition is achieved as evidenced by 91+/92+ peak ratios in excess of 20 for dissociation of the n-butylbenzene molecular ion.1 The amount of internal energy can be controlled directly by varying the amplitude of the pulse. Dissociation efficiencies are less than those typically observed for collision-induced dissociation in the ion trap and range from 50% at low internal energy deposition to 5% at maximum internal energy deposition.