DETERMINATION OF X-RAY-SPECTRA AND OF THE SCATTERED COMPONENT UP TO 300 KV

被引:7
作者
LAITANO, RF [1 ]
PANI, R [1 ]
PELLEGRINI, R [1 ]
机构
[1] UNIV ROME LA SAPIENZA,DIPARTIMENTO MED SPERIMENTALE,SEZ FIS MED & SANITARIA,I-00185 ROME,ITALY
关键词
X-RAY SPECTROMETRY; REFERENCE RADIATIONS; MONTE CARLO PROCEDURE;
D O I
10.1118/1.596742
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Several x-ray spectra, including those of the ISO reference radiations, were measured by 2-hp Ge detectors. Measurements were carried out in different experimental conditions with regard to detector size, beam collimation, and SDD. A stripping procedure to improve the spectrum analysis was developed on the basis of a detailed evaluation (by means of a Monte Carlo method) of the detector's spurious effects. These effects include K-photon escape, Compton photon escape, electron escape, and collimation effect. The stripping procedure also allows us to determine directly the spectra of possible scattered radiation reaching the detector in addition to the primary beam. When the primary beam is heavily filtered, the leakage radiation from the x-ray tube housing scattered onto the detector may not be negligible even when the x-ray tube is provided with appreciable shielding. Possible practical consequences of these effects are discussed. The results obtained on the ISO x-ray spectra are in agreement with previous determinations. The advantage of the present procedure is its more immediate applicability to Ge detectors of any size and with different beam collimation conditions.
引用
收藏
页码:934 / 938
页数:5
相关论文
共 5 条
[1]   EFFECT OF COLLIMATORS ON THE MEASUREMENT OF DIAGNOSTIC-X-RAY SPECTRA [J].
KODERA, Y ;
CHAN, HP ;
DOI, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (07) :841-852
[2]   DIAGNOSTIC-X-RAY SPECTRA MEASUREMENTS USING A SILICON SURFACE-BARRIER DETECTOR [J].
PANI, R ;
LAITANO, RF ;
PELLEGRINI, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (09) :1135-1149
[3]   STRIPPING OF X-RAY BREMSSTRAHLUNG SPECTRA UP TO 300-KVP ON A DESK TYPE COMPUTER [J].
SEELENTAG, WW ;
PANZER, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (04) :767-780
[4]  
SEELENTAG WW, 1979, GSF BER, V560
[5]  
SELTZER SM, 1981, NUCL INSTRUM METHODS, V188, P133, DOI 10.1016/0029-554X(81)90332-3