Polymicrobial gram negative bacteremia associated with saline solution flush used with a needleless intravenous system

被引:20
作者
Chodoff, A [1 ]
Pettis, AM [1 ]
Schoonmaker, D [1 ]
Shelly, MA [1 ]
机构
[1] HIGHLAND HOSP,HOSP EPIDEMIOL TEAM,ROCHESTER,NY 14620
关键词
D O I
10.1016/0196-6553(95)90266-X
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: During a 2-week period, seven cases of nosocomial polymicrobial gram-negative rod bacteremia occurred on a 39-bed medical and cardiac step-down unit. Combinations of Enterobacter cloacae (seven isolates), klebsiella pneumoniae (five isolates), and Citrobacter freundii (two isolates) were isolated from blood cultures. Methods: Concurrent and retrospective chart reviews were used to look for further cases and common exposures. Epidemiologic methods were used to refine determination of common exposure. Restriction enzyme DNA analysis was performed on the isolates. Results: Concurrent and retrospective chart reviews revealed four additional possible cases during the same period. All case patients were exposed, through peripheral saline solution locks or central venous catheters, to saline solution ''flush'' from a central 0.9% saline solution bag and a needleless dispensing pin. Epidemiologic methods implicated probable extrinsic contamination of a single bag and pin used during a 24- 48-hour period (Fisher's Exact Test, p < 0.002). There were no other common exposures. Restriction enzyme DNA analysis of the isolates further supported a common source for the outbreak. Conclusions: The introduction of needleless intravascular systems has been embraced for employee protection. Our report is the first to raise the question of patient safety with such systems. This outbreak highlights the inherent risks in rapid introduction of new technologies and points out the delicate balance among patient health, employee safety, and cost containment.
引用
收藏
页码:357 / 363
页数:7
相关论文
共 18 条
[1]  
Occupational Safety and Health Administration, Occupational exposure to bloodborne pathogens: final rule, Federal Register, 56, 235, pp. 64174-64182, (1991)
[2]  
Adams, Zehrer, Thomas, Comparison of a needleless system with conventional heparin locks, AJIC Am J Infect Control, 21, pp. 263-269, (1993)
[3]  
Gartner, Impact of a needleless intravenous system in a university hospital, AJIC Am J Infect Control, 20, pp. 75-79, (1992)
[4]  
Skolnick, LaRocca, Barba, Paicius, Evaluation and implementation of a needleless intravenous system: making needle sticks a needless problem, AJIC Am J Infect Control, 21, pp. 39-41, (1993)
[5]  
Dugleux, LeCoutour, Hecquard, Oblin, Septicemia caused by contaminated parenteral nutrition pouches: the refrigerator as an unusual cause, JPEN J Parenter Enter Nutr, 15, pp. 474-475, (1991)
[6]  
Jarvis, Highsmith, Allen, Haley, Polymicrobial bacteremia associated with lipid emulsion in a neonatal intensive care unit, Pediatr Infect Dis, 13, pp. 228-234, (1989)
[7]  
Koopman, Fighetti de Olave, An extensive outbreak of Klebsiella pneumoniae bacteremias from in-use contamination of IV, bottles, Bull Pan Am Health Organ, 14, pp. 185-192, (1980)
[8]  
Levy, Van Laethem, Verhaegen, Perpete, Butzler, Wenzel, Contaminated enteral nutrition solutions as a cause of nosocomial bloodstream infection: a study using plasmid fingerprinting, JPEN J Parenter Enter Nutr, 13, pp. 228-234, (1989)
[9]  
Maki, Epidemic nosocomial bacteremias, CRC handbook of hospital infections, pp. 430-431, (1981)
[10]  
Maki, Martin, Nationwide epidemic of septicemia caused by contaminated infusion products. IV. Growth of microbial pathogens in fluids for intravenous infusion, J Infect Dis, 131, pp. 267-272, (1975)