AN ELECTRON-DENSITY CALIBRATION PHANTOM FOR CT-BASED TREATMENT PLANNING COMPUTERS

被引:143
作者
CONSTANTINOU, C [1 ]
HARRINGTON, JC [1 ]
DEWERD, LA [1 ]
机构
[1] RADIAT MEASUREMENTS INC,MIDDLETON,WI 53562
关键词
TREATMENT PLANNING; ELECTRON DENSITY; COMPUTERIZED TOMOGRAPHY; QUALITY ASSURANCE; PHANTOM;
D O I
10.1118/1.596862
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In order to correct for tissue heterogeneities on a voxel-by-voxel basis during CT-based treatment planning, the relationship between the correction factor (CF) and the CT number in Hounsfield units (HU) for the scanner in use must be established. Since the relationship between CF and electron density (rho(e)) of various materials is well documented, the rho(e) vs HU is required for direct computation of the correction factors by treatment planning computers. A CT phantom with 18 different tissue substitutes has been used to establish the rho(e) vs HU relationship. A description of the phantom and its contents is given and the calibration of the CT function of the planning computer is discussed.
引用
收藏
页码:325 / 327
页数:3
相关论文
共 16 条
[1]  
BATHO H F, 1964, J Can Assoc Radiol, V15, P79
[2]   COMPTON SCATTER IMAGING OF TRANSVERSE SECTIONS - AN OVERALL APPRAISAL AND EVALUATION FOR RADIOTHERAPY PLANNING [J].
BATTISTA, JJ ;
BRONSKILL, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (01) :81-99
[3]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[4]  
CONSTANTINOU C, 1987, MED PHYS, V14, P487
[5]  
CONSTANTINOU C, 1988, International Journal of Radiation Oncology, Biology, Physics, V15, P244
[6]  
CONSTANTINOU C, 1978, THESIS LONDON U
[7]  
CUNNINGHAM JR, 1982, PROGR MED PHYSICS, V1
[8]   USE OF CT SCANNERS IN MEGAVOLTAGE PHOTON-BEAM THERAPY PLANNING [J].
GEISE, RA ;
MCCULLOUGH, EC .
RADIOLOGY, 1977, 124 (01) :133-141
[9]   LUNG DOSE CORRECTIONS FOR 6-MV AND 15-MV X-RAYS [J].
MACKIE, TR ;
ELKHATIB, E ;
BATTISTA, J ;
SCRIMGER, J ;
VANDYK, J ;
CUNNINGHAM, JR .
MEDICAL PHYSICS, 1985, 12 (03) :327-332
[10]   ACCEPTANCE TESTING COMPUTERIZED RADIATION-THERAPY TREATMENT PLANNING SYSTEMS - DIRECT UTILIZATION OF CT SCAN DATA [J].
MCCULLOUGH, EC ;
HOLMES, TW .
MEDICAL PHYSICS, 1985, 12 (02) :237-242