QUANTITATIVE SPECT BY ATTENUATION CORRECTION OF THE PROJECTION SET USING TRANSMISSION DATA - EVALUATION OF A METHOD

被引:22
作者
ALMQUIST, H
PALMER, J
LJUNGBERG, M
WOLLMER, P
STRAND, SE
JONSON, B
机构
[1] UNIV LUND, DEPT CLIN PHYSIOL, S-22185 LUND, SWEDEN
[2] UNIV LUND, DEPT RADIAT PHYS, S-22185 LUND, SWEDEN
来源
EUROPEAN JOURNAL OF NUCLEAR MEDICINE | 1990年 / 16卷 / 8-10期
关键词
Attenuation correction; Build-up; Quantitative SPECT; Transmission measurement;
D O I
10.1007/BF00998154
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method for measuring attenuation coefficients in single-photon emission tomography (SPECT) is described and evaluated, together with a method for attenuation correction using these measured attenuation coefficients. Build-up, caused by scattered photons, is corrected for by a simple substitution in the algorithms. Transmission studies are performed with a99mTc- or57Co flood source, and emission phantom studies with99mTc line sources. The method is evaluated with variable but well-defined phantoms. The result is accurate attenuation coefficients for different densities, dimensions and geometries, and an accuracy of corrected emission activities of better than ∓ 10% in most cases. The present limitations of the method for attenuation correction are discussed. © 1990 Springer-Verlag.
引用
收藏
页码:587 / 594
页数:8
相关论文
共 36 条
  • [1] STEADY STATE MEASUREMENT OF REGIONAL VENTILATION TO PERFUSION RATIOS IN NORMAL MAN
    ANTHONISEN, NR
    DOLOVICH, MB
    BATES, DV
    [J]. JOURNAL OF CLINICAL INVESTIGATION, 1966, 45 (08) : 1349 - +
  • [2] AXELSSON B, 1984, J NUCL MED, V25, P490
  • [3] BAILEY DL, 1987, J NUCL MED, V28, P1925
  • [4] BAILEY DL, 1987, J NUCL MED, V28, P844
  • [5] Berger MJ, 1987, NBSIR873597
  • [6] BEDSIDE DETERMINATION OF EXTRA-VASCULAR LUNG WATER - NON-INVASIVE DOUBLE INDICATOR TECHNIQUE USING ANTIPYRINE-I-123, IN-113M-TRANSFERRIN AND EXTERNAL COUNTING
    BINSWANGER, RO
    ROSLER, H
    NOELPP, U
    MATTER, L
    HAERTEL, M
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1978, 3 (02): : 109 - 114
  • [7] ASSESSMENT OF REGIONAL VENTILATION BY CONTINUOUS INHALATION OF RADIOACTIVE KRYPTON-81M
    FAZIO, F
    JONES, T
    [J]. BRITISH MEDICAL JOURNAL, 1975, 3 (5985) : 673 - 676
  • [8] NON-INVASIVE ANALYSIS OF METABOLIC REACTIONS IN BODY-TISSUES, THE CASE OF MYOCARDIAL FATTY-ACIDS
    FEINENDEGEN, LE
    VYSKA, K
    FREUNDLIEB, C
    HOCK, A
    MACHULLA, HJ
    KLOSTER, G
    STOCKLIN, G
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1981, 6 (05): : 191 - 200
  • [9] FLOYD CE, 1985, J NUCL MED, V26, P403
  • [10] FLOYD CE, 1986, J NUCL MED, V27, P1577