FLEXIBLE MINIATURE RIBBON CABLES FOR LONG-TERM CONNECTION TO IMPLANTABLE SENSORS

被引:12
作者
HETKE, JF
NAJAFI, K
WISE, KD
机构
[1] Center for Integrated Sensors and Circuits, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor
基金
美国国家卫生研究院;
关键词
D O I
10.1016/0924-4247(90)87076-U
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes the development of miniature, multi-lead ribbon cables for connecting implantable sensors to host electronics, either implanted or in the external world. The cables are flexible, electrically stable and biocompatible. Two technologies have been developed: one using polyimide (DuPont PI-2555) and the other using silicon as the support material. The present cables are less than 200 μm wide, are 10 to 20 μm thick and can be of arbitrary length (typically 1-2 cm). They can support as many leads as desired and can be bonded to the sensor substrate using flip-chip bonding techniques. © 1990.
引用
收藏
页码:999 / 1002
页数:4
相关论文
共 6 条
[1]   SOLID-STATE ELECTRODES FOR MULTICHANNEL MULTIPLEXED INTRACORTICAL NEURONAL RECORDING [J].
BEMENT, SL ;
WISE, KD ;
ANDERSON, DJ ;
NAJAFI, K ;
DRAKE, KL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :230-241
[2]  
DOANE LM, 1988, J ELECTROCHEM SOC, V133, P3155
[3]   HISTOLOGICAL REACTION TO POLYIMIDE FILMS IN THE COCHLEA [J].
HAGGERTY, HS ;
LUSTED, HS .
ACTA OTO-LARYNGOLOGICA, 1989, 107 (1-2) :13-22
[4]   A HIGH-YIELD IC-COMPATIBLE MULTICHANNEL RECORDING ARRAY [J].
NAJAFI, K ;
WISE, KD ;
MOCHIZUKI, T .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1985, 32 (07) :1206-1211
[5]  
NAJAFI K, 1987, 4TH INT C SOL STAT S, P65
[6]  
WISE KD, 1989, VLSI MED, pCH10