Analysis of bearing surfaces of explanted cups can help to determine the wear mechanisms that are responsible for generation of wear debris. In this study a microscope polariscope was used to detect residual subsurface shear strains, deformation and subsurface cracks in explanted Charnley acetabular cups. The wear surfaces were compared to an acetabular cup from a hip joint simulator test. The six explanted cups that were studied had all failed after long periods of implantation, with penetrations ranging from 2.1 to 3.8 mm. The explanted and simulator cups both had a smooth, high-wear region. High residual subsurface shear strains were found in the high-wear region of most cups, with certain cups possessing subsurface cracks running parallel with the surface 5-10 mum deep, close to the areas of high residual subsurface shear strain. This was caused by plastic deformation and subsurface fatigue of the polymer surface.
引用
收藏
页码:52 / 57
页数:6
相关论文
共 15 条
[1]   LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .2. A MICROSCOPICAL STUDY OF THE SURFACES OF CHARNLEY POLYETHYLENE ACETABULAR SOCKETS [J].
ATKINSON, JR ;
DOWSON, D ;
ISAAC, GH ;
WROBLEWSI, BM .
WEAR, 1985, 104 (03) :217-224
[2]   BIREFRINGENT STUDIES OF POLYETHYLENE WEAR SPECIMENS AND ACETABULAR CUPS [J].
COOPER, JR ;
DOWSON, D ;
FISHER, J .
WEAR, 1991, 151 (02) :391-402
[3]  
COOPER JR, 1992, 18TH P LEEDS LYON S, P29
[4]   CHARACTERISTICS OF ACETABULAR CUPS WORN IN HUMAN-BODY [J].
DOWLING, JM ;
ATKINSON, JR ;
DOWSON, D ;
CHARNLEY, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :375-382
[5]  
DOWSON D, 1985, P AM CHEM SOC, V287, P172
[6]  
FISHER J, 1992, J ENG MED H, V205, P73
[7]   APPLICATION OF TOPOGRAPHICAL ANALYSIS TO WEAR OF POLYMERS [J].
HOLLANDER, AE ;
LANCASTER, JK .
WEAR, 1973, 25 (02) :155-170
[8]   A RAT MODEL OF RESORPTION OF BONE AT THE CEMENT-BONE INTERFACE IN THE PRESENCE OF POLYETHYLENE WEAR PARTICLES [J].
HOWIE, DW ;
VERNONROBERTS, B ;
OAKESHOTT, R ;
MANTHEY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (02) :257-263
[9]  
Isaac G H, 1987, Eng Med, V16, P167
[10]  
JOBBINS B, 1988, ENG MED, V17, P111