THE EFFECT OF FREQUENCY AND MEAN AIRWAY PRESSURE ON VOLUME DELIVERY DURING HIGH-FREQUENCY OSCILLATION

被引:47
作者
CHAN, V
GREENOUGH, A
MILNER, AD
机构
[1] UNIV LONDON KINGS COLL HOSP,DEPT CHILD HLTH,LONDON SE5 9RS,ENGLAND
[2] UNIV LONDON KINGS COLL HOSP,DEPT PAEDIAT,LONDON SE5 9RS,ENGLAND
关键词
INVITRO AND INVIVO ASSESSMENT;
D O I
10.1002/ppul.1950150310
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The performance of a commercially available oscillator (SensorMedics 3100) at different frequencies was assessed. A frequency response curve of a pneumotachograph system was constructed and this was used to measure the volume delivered by the oscillator to a lung model. The volume delivered by a constant diaphragm displacement was demonstrated to be inversely proportional to the frequency, but unaffected by increasing mean airway pressure from 15 to 25 cm H2O. The volume delivered during high frequency oscillation (HFO) was then asessed in 8 infants, median gestational age 29 weeks. The infants were studied at two frequencies, 10 and 15 Hz, both of which were used at two levels of mean airway pressure (MAP): 2 and 5 cm H2O above the MAP level previously used during conventional ventilation. The delivered volume was not significantly different at the two MAP levels, but was significantly greater at 10 than 15 Hz at both MAP levels (P < 0.03); at MAP +2 cm H2O above baseline the reduction in delivered volume was from a median of 1.54 mL/kg (range, 0.88-3.12) at 10 Hz to 1.18 mL/kg (range, 0.65-4.5) at 15 Hz. These results suggest that higher frequencies would require an increase in the oscillator displacement if effective-gas exchange is to be maintained.
引用
收藏
页码:183 / 186
页数:4
相关论文
共 8 条
[1]  
CLARK RH, 1992, PEDIATRICS, V89, P5
[2]  
DORKIN HL, 1981, PEDIATR RES, V15, P717, DOI 10.1203/00006450-198104001-01664
[3]   FACTORS INFLUENCING MECHANICAL PERFORMANCE OF NEONATAL HIGH-FREQUENCY VENTILATORS [J].
FREDBERG, JJ ;
GLASS, GM ;
BOYNTON, BR ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (06) :2485-2490
[4]   PERFORMANCE OF RESPIRATORS AT FAST RATES COMMONLY USED IN NEONATAL INTENSIVE-CARE UNITS [J].
GREENOUGH, A ;
GREENALL, F .
PEDIATRIC PULMONOLOGY, 1987, 3 (05) :357-361
[5]   COMBINED CONVENTIONAL VENTILATION WITH HIGH-FREQUENCY OSCILLATION IN NEONATES [J].
HOSKYNS, EW ;
MILNER, AD ;
HOPKIN, IE .
EUROPEAN JOURNAL OF PEDIATRICS, 1991, 150 (05) :357-361
[6]   ALVEOLAR DATA IN HEALTHY, AWAKE NEONATES DURING SPONTANEOUS VENTILATION - A PRELIMINARY INVESTIGATION [J].
LAGNEAUX, D ;
MOSSAY, C ;
GEUBELLE, F ;
CHRISTIAENS, G .
PEDIATRIC PULMONOLOGY, 1988, 5 (04) :225-231
[7]   THE EFFECT OF LUNG-MECHANICS ON GAS-TRANSPORT DURING HIGH-FREQUENCY OSCILLATION [J].
SCHINDLER, M ;
SEEAR, M .
PEDIATRIC PULMONOLOGY, 1991, 11 (04) :335-339
[8]  
1990, J PEDIATR, V177, P939