SIMULTANEOUS ASSAYS OF GLUCOSE, URATE, AND CHOLESTEROL IN BLOOD-SERUM BY AMPEROMETRIC FLOW-INJECTION ANALYSIS

被引:20
作者
YAO, T
SATOMURA, M
NAKAHARA, T
机构
[1] Department of Applied Chemistry, College of Engineering, University of Osaka Prefecture, Sakai, Osaka, 593, Gakuencho
关键词
ENZYME REACTOR; FLOW INJECTION ANALYSIS; SIMULTANEOUS DETERMINATION; GLUCOSE; URIC ACID; CHOLESTEROL;
D O I
10.1002/elan.1140070209
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Glucose, uric acid, and cholesterol in serum were assayed simultaneously in a how injection system by using a 16-way switching valve and a parallel configuration of three immobilized enzyme reactors. Nafion film-coated and Nafion/cellulose acetate him-coated platinum electrodes were used to selectively detect the hydrogen peroxide generated enzymatically in the enzyme reactors, without any interferences from electroactive species such as L-ascorbate and proteins present in the serum. The signal current was linearly related to the glucose concentration in the range of 5-700 mg/dL, to the urate concentration in the range of 1-80 mg/dL, and to the cholesterol concentration in the range of 5-100 mg/dL. The precision (coefficient of variation) was better than 2.8% for the simultaneous assay of the three species in human control serum. The assay speed was up to 32 samples per hour.
引用
收藏
页码:143 / 146
页数:4
相关论文
共 21 条
[1]   PULSED AMPEROMETRIC DETECTION OF GLUCOSE IN BIOLOGICAL-FLUIDS AT A SURFACE-MODIFIED GOLD ELECTRODE [J].
BINDRA, DS ;
WILSON, GS .
ANALYTICAL CHEMISTRY, 1989, 61 (22) :2566-2570
[2]  
Guilbault G., 1984, ANAL USES IMMOBILIZE
[3]   CHARACTERIZATION OF PERFLUOROSULFONIC ACID POLYMER COATED ENZYME ELECTRODES AND A MINIATURIZED INTEGRATED POTENTIOSTAT FOR GLUCOSE ANALYSIS IN WHOLE-BLOOD [J].
HARRISON, DJ ;
TURNER, RFB ;
BALTES, HP .
ANALYTICAL CHEMISTRY, 1988, 60 (19) :2002-2007
[4]   A THIN-LAYER ELECTROCHEMICAL DETECTOR COATED WITH NAFION FILM FOR LIQUID-CHROMATOGRAPHY [J].
JI, HM ;
WANG, EK .
JOURNAL OF CHROMATOGRAPHY, 1987, 410 (01) :111-120
[5]  
MARKOVARGA G, 1992, ELECTROANAL, V4, P1
[6]   SIMULTANEOUS DETERMINATION OF GLUCOSE, ETHANOL AND LACTATE IN ALCOHOLIC BEVERAGES AND SERUM BY AMPEROMETRIC FLOW-INJECTION ANALYSIS WITH IMMOBILIZED ENZYME REACTORS [J].
MATSUMOTO, K ;
MATSUBARA, H ;
HAMADA, M ;
UKEDA, H ;
OSAJIMA, Y .
JOURNAL OF BIOTECHNOLOGY, 1990, 14 (01) :115-126
[7]   ION-EXCHANGE AND TRANSPORT OF NEUROTRANSMITTERS IN NAFION FILMS ON CONVENTIONAL AND MICROELECTRODE SURFACES [J].
NAGY, G ;
GERHARDT, GA ;
OKE, AF ;
RICE, ME ;
ADAMS, RN ;
MOORE, RB ;
SZENTIRMAY, MN ;
MARTIN, CR .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1985, 188 (1-2) :85-94
[8]   ELECTROPOLYMERIZED 1,2-DIAMINOBENZENE AS A MEANS TO PREVENT INTERFERENCES AND FOULING AND TO STABILIZE IMMOBILIZED ENZYME IN ELECTROCHEMICAL BIOSENSORS [J].
SASSO, SV ;
PIERCE, RJ ;
WALLA, R ;
YACYNYCH, AM .
ANALYTICAL CHEMISTRY, 1990, 62 (11) :1111-1117
[9]   FLOW-INJECTION ANALYSIS (FIA) BASED ON ENZYMES OR ANTIBODIES - APPLICATIONS IN THE LIFE SCIENCES [J].
SCHMID, RD ;
KUNNECKE, W .
JOURNAL OF BIOTECHNOLOGY, 1990, 14 (01) :3-31
[10]   SURFACE-MODIFIED ELECTROCHEMICAL DETECTOR FOR LIQUID-CHROMATOGRAPHY [J].
SITTAMPALAM, G ;
WILSON, GS .
ANALYTICAL CHEMISTRY, 1983, 55 (09) :1608-1610