摘要
A micropower signal-processor IC is the key component of an implantable telemetry system for neurophysiology (TSN). The bipolar/JFET/I**2 L chip uses digital and low-noise analog circuitry to amplify, filter, and multiplex eight channels of neural, electrogram, and temperature data from unanesthetized and freely moving animals. Fully integrated continuous-time bandpass amplifiers incorporate a frequency-sensitive feedback network to prevent the amplification of input offset voltage. The system can telemeter data for over 500 h, permitting long-term neurophysiological investigations.
引用
收藏
页码:1185 / 1193
页数:9
相关论文
共 24 条
[1]  
ALLEN HV, 1979, BIOTELEM PAT MON, V6, P147
[2]  
ALLEN HV, 1981, FEB ISSCC, P166
[3]   PRECISION DIFFERENTIAL VOLTAGE-CURRENT CONVERTER [J].
CAPRIO, R .
ELECTRONICS LETTERS, 1973, 9 (06) :147-148
[4]  
CROMWELL L, 1973, BIOMEDICAL INSTRUMEN, P249
[5]  
DORMAN MG, 1985, THESIS STANFORD U ST, P217
[6]   A BIPOLAR VOLTAGE-CONTROLLED TUNABLE FILTER [J].
FUKAHORI, K .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1981, 16 (06) :729-737
[7]  
GLASER AB, 1977, INTEGRATED CIRCUIT E, P22
[8]  
Heller H. C., 1978, STRATEGIES COLD, V1978, P225, DOI [10.1016/B978-0-12-734550-5. 50012-0, DOI 10.1016/B978-0-12-734550-5.50012-0]
[9]  
HSIEH KC, 1981, IEEE J SOLID-ST CIRC, V16, P708
[10]  
KLIGN JAJ, 1978, BIOTELEMETRY, V4, P45