ASSESSMENT OF THE RESPIRATORY METABOLISM IN THE SKIN FROM TRANSCUTANEOUS MEASUREMENTS OF PO2 AND PCO2 - POTENTIAL FOR NONINVASIVE MONITORING OF RESPONSE TO TUBERCULIN SKIN TESTING

被引:21
作者
ABBOT, NC
SPENCE, VA
BECK, JS
CARNOCHAN, FMT
GIBBS, JH
LOWE, JG
机构
[1] UNIV DUNDEE,NINEWELLS HOSP & MED SCH,DEPT PATHOL,DUNDEE DD1 9SY,SCOTLAND
[2] UNIV DUNDEE,NINEWELLS HOSP & MED SCH,VASC LAB,DUNDEE DD1 9SY,SCOTLAND
来源
TUBERCLE | 1990年 / 71卷 / 01期
关键词
D O I
10.1016/0041-3879(90)90055-D
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
A method is described for non-invasive transcutaneous (tc) measurement of tissue respiratory gas tensions in the skin on the forearm for study of delayed hypersensitivity reactions in man. Steady state values for tcpO2 and tcpCO2 were measured, and the skin respiratory rate (oxygen consumption) and the tissue pH were estimated from the changes in tcpO2 and tcpCO2 observed after interruption of the arterial circulation by cuff occlusion for 4 minutes. The extent of within-experiment and between subject variation in the steady-state measurements was not great (coefficient of variation 10%): tcpCO2.ss (steady state) was higher in men and tcpO2.ss was higher in women, but the extent of these sex differences was also small. Reference ranges have been established for tc measurements and calculated indices of tissue respiration in the undisturbed forearm skin of normal volunteers, against which the changes induced by tuberculin testing can be assessed. Severe changes, indicative of profound hypoxia and acidosis, are seen in intense delayed hypersensitivity reactions. Similar, but less severe changes were seen at the site of skin tests on BCG-vaccinated subjects who were 'negative' by conventional criteria of measurement of dermal induration and they became greatly exaggerated after successful re-vaccination. Intradermal injection of saline did not induce hypoxia or local acidosis. These new methods are very sensitive indicators of the tissue response in the DHS reaction. © 1990.
引用
收藏
页码:15 / 22
页数:8
相关论文
共 25 条
[1]  
BECK JS, 1988, BRIT J EXP PATHOL, V69, P771
[2]  
BECK JS, 1986, CLIN EXP IMMUNOL, V64, P484
[3]  
BECK JS, 1986, J IMMUNOL METHODS, V86, P125, DOI 10.1016/0022-1759(86)90275-9
[4]  
BECK JS, 1986, IMMUNOLOGY, V58, P209
[5]   STATISTICS ON MICROCOMPUTERS - A NON-ALGEBRAIC GUIDE TO THE APPROPRIATE USE OF STATISTICAL PACKAGES IN BIOMEDICAL-RESEARCH AND PATHOLOGY LABORATORY PRACTICE - 6 STATISTICAL-METHODS FOR DIAGNOSTIC-TESTS [J].
BROWN, RA ;
BECK, JS .
JOURNAL OF CLINICAL PATHOLOGY, 1989, 42 (03) :225-230
[6]  
CARNOCHAN F, IN PRESS AGENTS ACTI
[7]   REPRODUCIBILITY OF TCPO2 MEASUREMENTS IN NORMAL VOLUNTEERS [J].
COLEMAN, LS ;
DOWD, GSE ;
BENTLEY, G .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1986, 7 (03) :259-263
[8]  
Diem K., 1962, DOCUMENTA GEIGY SCI
[9]  
GOTHGEN I, 1978, ACTA ANAESTHESIOL S, V67, P66
[10]  
HARRISON DK, 1986, IMMUNOLOGY, V59, P497