COMPARISON OF 2 DIGESTION METHODS FOR THE DETERMINATION OF SELENIUM IN BIOLOGICAL SAMPLES

被引:36
作者
DUCROS, V
RUFFIEUX, D
BELIN, N
FAVIER, A
机构
关键词
SELENIUM DETERMINATION; BIOLOGICAL SAMPLE; WET-ASHING; MICROWAVE DIGESTION;
D O I
10.1039/an9941901715
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two digestion methods for the determination of selenium in biological samples have been compared: a classic digestion in a heating block by means of wet acid-ashing (HNO3-HClO4) and a microwave digestion in an open-vessel system by heating under reflux with HNO3-H2O2. The two methods were used to determine Se in plasma samples and different reference materials. There was no difference between the mean results, and the correlation coefficient for wet acid-ashing versus microwave digestion was 0.94. The two methods yielded accurate results for the certified materials analysed.
引用
收藏
页码:1715 / 1717
页数:3