HADAMARD imaging spectrometer with micro slit matrix

被引:26
作者
Riesenberg, R [1 ]
Dillner, U [1 ]
机构
[1] Inst Phys High Technol, D-07702 Jena, Germany
来源
IMAGING SPECTROMETRY V | 1999年 / 3753卷
关键词
Imaging spectrometer; HADAMARD spectrometer; digital imaging; spatial light modulator;
D O I
10.1117/12.366283
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
The concept for an imaging spectral sensor is presented. The starting point is an advanced raster scanning with multiplexing in spatial terms.' The essential components are an entrance array, an imaging grating and a detector array - a double array architecture(1). A programmable micro slit matrix is used as a two-dimensional HADAMARD-mask. It scans the two-dimensional images to be processed in a one-dimensional spectrometer. A multiplexing in spatial terms and simultaneously in spectral terms is possible. The spectra recorded by a commercial detector array spectrometer. At the present state the multiplexing of more than 100 spatial pixels within the quality of spectrometry is demonstrated. The concept provides a HADAMARD or multiplex advantage resulting from the micro slit matrix (increase of SMR by root n/2, n-number of imaging pixels) and a multidetector advantage resulting from the linear detector array of the spectrometer (increase of SNR by root N, N-number of detector elements). Experiments reported on a further paper.
引用
收藏
页码:203 / 213
页数:11
相关论文
共 13 条
[1]  
DERENIAK EL, COMMUNICATION
[2]  
DeVerse RA, 1998, AM LAB, V30, p112S
[3]  
HARVID M, 1979, HADAMARD TRANSFORM O
[4]  
Hornbeck L. J., 1995, DIGITAL LIGHT PROCES
[5]   Design of spatial light modulator microdevices - micro slit arrays [J].
Riesenberg, R ;
Seifert, T .
DESIGN, TEST, AND MICROFABRICATION OF MEMS AND MOEMS, PTS 1 AND 2, 1999, 3680 :406-414
[6]  
RIESENBERG R, 1997, PITTCON 97, P1059
[7]  
RIESENBERG R, 1997, 2 ROUND TABL MICR TE, P145
[8]  
RIESENBERG R, 1999, IN PRESS P SPIE
[9]  
RIESENBERG R, 1998, SPIES S REM SENS ATM
[10]  
RIESENBERG R, 1999, PITTCON 99, P581