SENSITIZED DETERMINATION OF SULFITE USING FLOW-INJECTION WITH CHEMILUMINESCENT DETECTION

被引:43
作者
PAULLS, DA [1 ]
TOWNSHEND, A [1 ]
机构
[1] UNIV HULL,SCH CHEM,KINGSTON HULL HU6 7RX,N HUMBERSIDE,ENGLAND
关键词
CHEMILUMINESCENCE; SULFITE; FLOW INJECTION; 3-CYCLOHEXYLAMINOPROPANESULFONIC ACID; CYCLOOCTYLAMINE;
D O I
10.1039/an9952000467
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The acidic oxidation of sulfite by Ce-IV is weakly chemiluminescent. A comparative study of the chemiluminescent determination of sulfite with the use of 3-cyclohexylaminopropanesulfonic acid (CAPS) and of cyclooctylamine as sensitizers of the reaction was made using a flow injection procedure. Despite a slight decrease in sensitivity; it was found that cyclooctylamine was superior to CAPS as a sensitizer and a detection limit of 5.4 x 10(-7) mol l(-1) sulfite was achieved compared with 1 x 10(-5) mol l(-1) using CAPS.
引用
收藏
页码:467 / 469
页数:3
相关论文
共 9 条
[1]   FLOW-INJECTION CHEMILUMINESCENCE DETERMINATION OF SULFITE [J].
ALTAMRAH, SA ;
TOWNSHEND, A ;
WHEATLEY, AR .
ANALYST, 1987, 112 (06) :883-886
[2]   CONTINUOUS-FLOW CHEMILUMINESCENCE DETERMINATION OF SOME CORTICOSTEROIDS [J].
KOUKLI, II ;
CALOKERINOS, AC .
ANALYST, 1990, 115 (12) :1553-1557
[3]   EFFECT OF SENSITIZERS ON THE CHEMILUMINESCENT REDUCTION OF CERIUM(IV) BY SULFITE [J].
KOUKLI, II ;
SARANTONIS, EG ;
CALOKERINOS, AC .
ANALYTICAL LETTERS, 1990, 23 (07) :1167-1179
[4]   FLOW-INJECTION CHEMILUMINOMETRIC DETERMINATION OF SODIUM CYCLAMATE [J].
PSARELLIS, IM ;
SARANTONIS, EG ;
CALOKERINOS, AC .
ANALYTICA CHIMICA ACTA, 1993, 272 (02) :265-270
[5]  
STAUFF J, 1978, Z NATURFORSCH B, V33, P293
[6]   CHEMILUMINESCENCE TECHNIQUE FOR MEASURING ATMOSPHERIC TRACE CONCENTRATIONS OF SULFUR-DIOXIDE [J].
STAUFF, J ;
JAESCHKE, W .
ATMOSPHERIC ENVIRONMENT, 1975, 9 (11) :1038-1039
[7]   FLOW-INJECTION CHEMILUMINOMETRIC ANALYSIS OF SOME STEROIDS BY THEIR SENSITIZING EFFECT ON THE BROMATE SULFITE REACTION [J].
SYROPOULOS, AB ;
SARANTONIS, EG ;
CALORKERINOS, AC .
ANALYTICA CHIMICA ACTA, 1990, 239 (02) :195-202
[8]  
WHEATLEY AR, 1983, THESIS U HULL
[9]   THE DETERMINATION OF SULFITE IN A FLOW-INJECTION SYSTEM WITH CHEMI-LUMINESCENCE DETECTION [J].
YAMADA, M ;
NAKADA, T ;
SUZUKI, S .
ANALYTICA CHIMICA ACTA, 1983, 147 (MAR) :401-404