PARALLEL CRYOGENIC TRAPPING MULTIDIMENSIONAL GAS-CHROMATOGRAPHY WITH DIRECTLY LINKED INFRARED AND MASS-SPECTRAL DETECTION

被引:19
作者
KROCK, KA [1 ]
RAGUNATHAN, N [1 ]
WILKINS, CL [1 ]
机构
[1] UNIV CALIF RIVERSIDE, DEPT CHEM, RIVERSIDE, CA 92521 USA
来源
JOURNAL OF CHROMATOGRAPHY | 1993年 / 645卷 / 01期
关键词
D O I
10.1016/0021-9673(93)80629-M
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes successful implementation of multiple trap multidimensional gas chromatography (GC) directly linked with parallel Fourier transform infrared (FT-IR) and mass spectroscopy (MS) detectors. For complex mixture separation, six parallel cryogenic traps are interposed between a first-stage GC pre-column and a second-stage GC analytical column. With proper choices of heart cut time intervals, this multiple parallel cryogenic trap approach allows expansion of the practical analytical dynamic range and accommodates the separation constraints of a combined GC-FT-IR-MS system. An unleaded gasoline sample is analyzed with respect to the degree of secondary separation and to the reproducibility of adjusted retention times. The results of cryogenically trapping heart cuts using both with 72- and 12-s trapping times are presented. Shortening the trapping times from 72 to 12 s results in a significant improvement in the analytical column separation and identification with infrared and mass spectra. Thus, use of multidimensional separation coupled with multispectral detection may provide increased reliability of complex mixture component identification.
引用
收藏
页码:153 / 159
页数:7
相关论文
共 19 条
[1]  
BERTSCH W, 1990, CHROMATOGRAPHIC SCI, V50, P74
[2]   DETERMINATION OF SULFURATED COMPOUNDS IN TAGETES-PATULA CV NANA ESSENTIAL OIL BY GAS-CHROMATOGRAPHY WITH MASS-SPECTROMETRIC, FOURIER-TRANSFORM INFRARED AND ATOMIC EMISSION SPECTROMETRIC DETECTION [J].
BICCHI, C ;
FRATTINI, C ;
PELLEGRINO, G ;
RUBIOLO, P ;
RAVERDINO, V ;
TSOUPRAS, G .
JOURNAL OF CHROMATOGRAPHY, 1992, 609 (1-2) :305-313
[3]   GAS-CHROMATOGRAPHY FOURIER-TRANSFORM INFRARED MASS-SPECTROMETRY USING A MASS SELECTIVE DETECTOR [J].
COOPER, JR ;
BOWATER, IC ;
WILKINS, CL .
ANALYTICAL CHEMISTRY, 1986, 58 (13) :2791-2796
[4]   UTILIZATION OF SPECTROMETRIC INFORMATION IN LINKED GAS-CHROMATOGRAPHY FOURIER-TRANSFORM INFRARED-SPECTROSCOPY MASS-SPECTROMETRY [J].
COOPER, JR ;
WILKINS, CL .
ANALYTICAL CHEMISTRY, 1989, 61 (14) :1571-1577
[5]  
DUNCAN WP, 1989, CLIN CHEM, V35, P1279
[6]  
GIDDINGS JC, 1990, CHROMATOGRAPHIC SCI, V50, P15
[7]   ANALYSIS OF FLUE-CURED TOBACCO ESSENTIAL OIL BY HYPHENATED ANALYTICAL TECHNIQUES [J].
GORDON, BM ;
UHRIG, MS ;
BORGERDING, MF ;
CHUNG, HL ;
COLEMAN, WM ;
ELDER, JF ;
GILES, JA ;
MOORE, DS ;
RIX, CE ;
WHITE, EL .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1988, 26 (04) :174-180
[8]   RECONSTRUCTION OF GAS CHROMATOGRAMS FROM INTERFEROMETRIC GAS CHROMATOGRAPHY INFRARED SPECTROMETRY DATA [J].
HASETH, JAD ;
ISENHOUR, TL .
ANALYTICAL CHEMISTRY, 1977, 49 (13) :1977-1981
[9]   COMPONENT ANALYSIS OF EUCALYPTUS OIL BY GAS-CHROMATOGRAPHY FOURIER TRANSFORM-INFRARED SPECTROMETRY MASS-SPECTROMETRY [J].
HEDGES, LM ;
WILKINS, CL .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (08) :345-350
[10]   ISOMERIC CHARACTERIZATION OF POLYCHLORINATED-BIPHENYLS USING GAS-CHROMATOGRAPHY FOURIER-TRANSFORM INFRARED GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
HEMBREE, DM ;
SMYRL, NR ;
DAVIS, WE ;
WILLIAMS, DM .
ANALYST, 1993, 118 (03) :249-251