COMPLICATIONS RELATED TO MODULARITY OF TOTAL HIP COMPONENTS

被引:76
作者
BARRACK, RL
BURKE, DW
COOK, SD
SKINNER, HB
HARRIS, WH
机构
[1] MASSACHUSETTS GEN HOSP,ORTHOPAED BIOMECH LAB,BOSTON,MA 02114
[2] UNIV CALIF SAN FRANCISCO,DEPT ORTHOPAED SURG,SAN FRANCISCO,CA 94143
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1993年 / 75卷 / 05期
关键词
D O I
10.1302/0301-620X.75B5.8376421
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report complications from the use of modular components in 20 hip replacements in 18 patients. Fifteen complications (in 13 patients) were related to failure of a modular interface after operation. Femoral head detachment from its trunnion was seen in 6 hips from trauma (3), reduction of a dislocation (2), and normal activity (1). In one case the base of the trunnion fractured below an extra-long modular head. In seven other hips the modular polyethylene liner dislodged from its shell, causing severe damage to the shell in four cases with extensive metallosis. In one other hip an asymmetrical polyethylene liner rotated, resulting in impingement of the femoral component and recurrent dislocation. Operative errors were seen in five cases: implantation of a trial acetabular component in one; and mismatching between the size of the femoral head and the acetabular component in the others. Surgeons who use hip replacements with modular components should be aware of the potential for operative error and of the importance of early treatment for postoperative mechanical failure.
引用
收藏
页码:688 / 692
页数:5
相关论文
共 12 条
[1]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[2]   DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES [J].
BRIEN, WW ;
SALVATI, EA ;
WRIGHT, TM ;
NELSON, CL ;
HUNGERFORD, DS ;
GILLIAM, DL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) :1548-1550
[3]   CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[4]  
COLLIER JP, 1990, CLIN ORTHOP RELAT R, V261, P107
[5]  
Ferenz C C, 1988, J Arthroplasty, V3, P201
[6]  
Harris W H, 1971, Clin Orthop Relat Res, V81, P105, DOI 10.1097/00003086-197111000-00016
[7]   FURTHER FOLLOW-UP ON SOCKET FIXATION USING A METAL-BACKED ACETABULAR COMPONENT FOR TOTAL HIP-REPLACEMENT - A MINIMUM 10-YEAR FOLLOW-UP-STUDY [J].
HARRIS, WH ;
PENENBERG, BL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (08) :1140-1143
[8]   DISASSEMBLY OF A MODULAR ACETABULAR COMPONENT OF A TOTAL HIP-REPLACEMENT ARTHROPLASTY - A CASE-REPORT [J].
KITZIGER, KJ ;
DELEE, JC ;
EVANS, JA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :621-623
[9]  
Krushell R J, 1991, J Arthroplasty, V6, P97, DOI 10.1016/S0883-5403(11)80002-1
[10]   FORMATION OF A FULMINANT SOFT-TISSUE PSEUDOTUMOR AFTER UNCEMENTED HIP-ARTHROPLASTY - A CASE-REPORT [J].
SVENSSON, O ;
MATHIESEN, EB ;
REINHOLT, FP ;
BLOMGREN, G .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (08) :1238-1242