A MICROPROCESSOR-REGULATED CONSTANT VOLTAGE, CURRENT, WATTAGE, AND TEMPERATURE ELECTROPHORESIS POWER-SUPPLY

被引:1
作者
MINCEY, DW
KUZIOR, KJ
ALLEN, LH
FREASE, JS
STRASSER, IN
机构
[1] Department of Chemistry, Youngstown State University, Youngstown
关键词
Analog to digital converters - Control voltages - Electrophoresis cell - End point detection - Feedback elements - Isoelectric focusing - Microcomputer system - Operating parameters;
D O I
10.1016/0003-2697(91)90003-C
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The analog control circuitry typically found in commercial electrophoresis power supplies was replaced by a digital microcomputer. Analog to digital converters were used to monitor the voltage applied to and current passed through an electrophoresis cell. Microcomputer programming was employed to compare converter input values with preselected operating parameters and then calculate a required output voltage. Timing sequences were generated through programming utilizing clocks located on the interface boards. A digital to analog converter was employed to apply a control voltage to a constant voltage power supply. This process was completed at least 20 times each second. BASIC programming subroutines were written to maintain constant voltage, current, power (wattage), and temperature. To these operating procedures, other techniques such as automated endpoint detection of isoelectric focusing and pulsed waveform outputs were easily added. This power supply containing a microcomputer system as the feedback element was shown to have a greater stability and versatility than conventional supplies. © 1991.
引用
收藏
页码:168 / 172
页数:5
相关论文
共 9 条
[1]  
ANDREWS TH, 1981, ELECTROPHORESIS THEO
[2]  
BIER M, 1959, ELECTROPHORESIS
[3]  
GORMAN WW, 1984, ACPR APR, P35
[4]  
JEPPSSON JO, 1977, LKB307 APPL NOT
[5]  
KAPLAN LA, 1984, CLIN CHEM
[6]   ISOELECTRIC-FOCUSING IN AGAROSE [J].
SARAVIS, CA ;
ZAMCHECK, N .
JOURNAL OF IMMUNOLOGICAL METHODS, 1979, 29 (01) :91-96
[7]   HIGHLY REGULATED, RECORDING CONSTANT POWER, VOLTAGE, AND CURRENT SUPPLY FOR ELECTROPHORESIS AND ISOELECTRIC-FOCUSING [J].
WILLIAMS, JM ;
CATSIMPOOLAS, N .
ANALYTICAL BIOCHEMISTRY, 1976, 71 (02) :555-563
[8]  
WINTER A, 1977, LKB250 APPL NOT
[9]  
1984, SERUM PROTEIN ELECTR, V1