CHEST RADIOGRAPHY - A SURVEY OF TECHNIQUES AND EXPOSURE LEVELS CURRENTLY USED

被引:19
作者
BUTLER, PF [1 ]
CONWAY, BJ [1 ]
SULEIMAN, OH [1 ]
KOUSTENIS, GH [1 ]
SHOWALTER, CK [1 ]
机构
[1] US FDA, CTR DEVICES & RADIOL HLTH, 5600 FISCHERS LANE, HFZ-240, ROCKVILLE, MD 20857 USA
关键词
D O I
10.1148/radiology.156.2.4011920
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In diagnostic radiology, the routine measurement of exposure levels for a reference patient were estimated is an important part of an effective quality assurance program. In the USA chest radiography is the most frequent examination and has the lowest exposure level of all radiologic examinations. The amount of exposure an average patient received from both manual and automatic exposure-controlled radiographic techniques [and personnel] by using a patient-equivalent chest phantom during measurements. A sensitometric procedure was used to assess processor performance. The mean exposure from 194 chest systems was 20 mR (5.16 .times. 10-5 C/kg); the mean film density, 1.38; and the mean processing speed, 108. A wide range of radiographic techniques, processing conditions and screen-film speeds are currently being used. With the information given in this study, investigators can begin to identify the problems that lead to unusual exposure levels and, perhaps, poor image quality.
引用
收藏
页码:533 / 536
页数:4
相关论文
共 10 条
  • [1] BUNGE RE, 1980, FDA US DEPT HHS PUBL, P172
  • [2] BUTLER PF, 1980, 14TH P MID HLTH PHYS, P241
  • [3] CLEARE HM, 1977, US DHEW FDA778036 PU, P11
  • [4] BEAM QUALITY INDEPENDENT ATTENUATION PHANTOM FOR ESTIMATING PATIENT EXPOSURE FROM X-RAY AUTOMATIC EXPOSURE CONTROLLED CHEST EXAMINATIONS
    CONWAY, BJ
    BUTLER, PF
    DUFF, JE
    FEWELL, TR
    GROSS, RE
    JENNINGS, RJ
    KOUSTENIS, GH
    MCCROHAN, JL
    RUETER, FG
    SHOWALTER, CK
    [J]. MEDICAL PHYSICS, 1984, 11 (06) : 827 - 832
  • [5] NOVITCH M, 1983, FED REGISTER, V48, P147
  • [6] PEARCE JG, 1979, P SPIE APPLICATIONS, V7, P173
  • [7] RADIOGRAPHIC FILM FOG IN THE DARKROOM
    SULEIMAN, OH
    SHOWALTER, CK
    GROSS, RE
    BUNGE, RE
    [J]. RADIOLOGY, 1984, 151 (01) : 237 - 238
  • [8] SULEMAN OH, 1982, FDA828189 US DEP HLT
  • [9] THOMPSON TT, 1979, CAHOONS FORMULATING, P99
  • [10] 1977, US DHEW FDA778006 PU