COMPUTER-ASSISTED PEDICLE SCREW FIXATION - A FEASIBILITY STUDY

被引:123
作者
AMIOT, LP
LABELLE, H
DEGUISE, JA
SATI, M
BRODEUR, P
RIVARD, CH
机构
[1] UNIV MONTREAL,HOP ST JUSTINE,DEPT ORTHOPED,MONTREAL,PQ H3T 1C5,CANADA
[2] ECOLE POLYTECH MONTREAL,INST GENIE BIOMED,DEPT GENIE MECAN,MONTREAL,PQ,CANADA
[3] ECOLE TECHNOL SUPER,MONTREAL,PQ,CANADA
关键词
COMPUTER GRAPHICS; PEDICLE SCREWS; SPINAL SURGERY;
D O I
10.1097/00007632-199505150-00019
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design, We evaluated a computer-assisted surgical tool for inserting pedicle screws. Objectives. This study reviewed the feasibility, usefulness, and accuracy of the proposed tool. Summary of Background Data, Reviews documented neurovascular damage caused by screw misplacement. Currently, screw hole position is assessed by radiologic means and curette palpation. Methods, Three sheep vertebrae and one artificial object were reconstructed three-dimensionally from computed tomography scan slices. At surgery, the surgeon's movements were displayed relative to the three-dimensional vertebrae on a computer screen. The tool was used to detect pedicles and to verify the position of drilled holes. In our laboratory, we calculated the system's accuracy by taking measurements on the artificial object. Results. All pedicles were identified with the computer. Five of the six drilled hole: positions were correctly represented. An accuracy of 4.5 mm +/- 1.1 mm RMS (root of the mean squared) and 1.6 degrees +/- 1.2 degrees were calculated. Conclusions, Results suggested the proposed system could be useful for pedicle detection and assessing the intravertebral location of a drilled hole. The proposed system could be used for many different orthopedic procedures where structures are hidden from the surgeon's view.
引用
收藏
页码:1208 / 1212
页数:5
相关论文
共 12 条
  • [1] Arun K.S., Huang T.S., Blostein S.D., Least-squares fitting of two 3-D sets, IEEE Transactions on Pattern Analysis, 9, pp. 698-700, (1987)
  • [2] Barnett G.W., Kormos D.W., Steiner C.P., Weisenberger J., Use of a frameiess, armless stereotaxic wand for brain tumor localization with 2-dimensional and 3-dimensional neuroimaging, Neurosurgery, 33, pp. 674-678, (1993)
  • [3] Beauchamp A., Dansereau J., De Guise J., Labelle H., Computer assisted digitization system for scoliotic spinal stereoradiographs, Proceedings of the International Symposium on 3D Scoliotic Deformities, pp. 18-25, (1992)
  • [4] Breau C., Shirazi-Adl A., De Guise J., Reconstruction of a human ligamentous lumbar spine using CT images a three- dimensional finite element mesh generation, Ann Biomed Eng, 19, pp. 291-302, (1991)
  • [5] Esses S.I., Sachs B.L., Dreyzin V., Complications Associated with the Technique of Pedicle Screw Fixation, Spine, 18, pp. 2231-2239, (1993)
  • [6] Evans A.C., Dai W., Collins L., Neelin P., Marrett S., Warping of a computerized 3D atlas to match brain image volumes for quantitative neuroanatomical and functional analysis. Pro-ceedings of the international society for optical engineering on Medical Imaging V: Image processing. San Jose, California: SPIE Press, 1445, pp. 236-246, (1991)
  • [7] Maciunas R.J., Galloway R.L., Fitzpatrick M.J., Mandave V., Edwards C.A., Allen G.S., A universal system for interactive image directed neurosurgery, Stereotact Funct Neurosurg, 58, pp. 108-113, (1992)
  • [8] Paul H.A., Mittlestadt B., Bargar W.L., Proceedings of the IEEE international conference on robotics and automation. Nice, France, A Surgical Robot for Total Hip Remplacement Surgery, (1992)
  • [9] Pelizzari C.A., Chen G., Spelbring D.R., Weichselbaum R.R., Chen C.T., Accurate three-dimensional registration of CT, PET or MR images of the brain, J Comput Assist Tomogr, 13, pp. 20-26, (1989)
  • [10] Roy-Camille R., Demeulanaere C., Osteosynthese du rachis dorsal, lombaire et lombo-sacre par plaques metalliques vissees dans Ies pedicules vertebraux et les apophyses articulaires, Presse Med, 78, pp. 1447-1448, (1970)