COMPARISON OF INTRAVASCULAR ULTRASONIC FINDINGS AFTER CORONARY BALLOON ANGIOPLASTY EVALUATED IN-VITRO WITH HISTOLOGY

被引:33
作者
VANDERLUGT, A
GUSSENHOVEN, EJ
STIJNEN, T
VANSTRIJEN, M
VANDRIEL, E
VANEGMOND, FC
SUYLEN, RJ
VANURK, H
机构
[1] ERASMUS UNIV ROTTERDAM,3000 DR ROTTERDAM,NETHERLANDS
[2] UNIV ROTTERDAM HOSP,ROTTERDAM,NETHERLANDS
关键词
D O I
10.1016/S0002-9149(99)80193-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study investigated whether vascular damage and quantitative changes observed with intravascular ultrasound at the most stenotic site are representative of the ultimate outcome after coronary balloon angioplasty. Atherosclerotic coronary arteries' (n = 40) were studied in vitro with intravascular ultrasound. From each vascular specimen, 10 corresponding intravascular ultrasound cross sections obtained before and after balloon angioplasty were selected for comparison with their histologic counterpart. Morphologic and quantitative data obtained from all cross sections were compared with data derived from the most stenotic site. The incidence of vascular damage (i.e., dissection, plaque rupture, and media rupture) at the most stenotic site was lower than that seen for each vascular specimen. The sensitivity of intravascular ultrasound in detecting these morphologic features for each vascular specimen was high for dissection and media rupture (79% and 76%, respectively) and low for plaque rupture (37%). After I balloon angioplasty, quantitative changes seen at the most stenotic site were greater than those in all cross sections: free lumen area +58% versus +29%, media-bound area +17% versus +12%, and plaque area reduction -9% versus -6%, respectively. The increase in free lumen area was caused predominantly by media-bound area increase (81%) and to a lesser extent by plaque area decrease (19%). This study revealed that a higher incidence of vascular damage is found when the whole segment is analyzed rather than 1 single cross section at the most stenotic site. Quantitative effects of coronary balloon angioplasty seen with intravascular ultrasound were greater at the most stenotic site than at all cross sections.
引用
收藏
页码:661 / 666
页数:6
相关论文
共 22 条
[1]   QUALITATIVE AND QUANTITATIVE CONTRASTS IN THE MECHANISMS OF LUMEN ENLARGEMENT BY CORONARY BALLOON ANGIOPLASTY AND DIRECTIONAL CORONARY ATHERECTOMY [J].
BRADEN, GA ;
HERRINGTON, DM ;
DOWNES, TR ;
KUTCHER, MA ;
LITTLE, WC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (01) :40-48
[2]  
CHIN AK, 1984, SURGERY, V95, P196
[3]   INVITRO VALIDATION OF 3-DIMENSIONAL INTRAVASCULAR ULTRASOUND FOR THE EVALUATION OF ARTERIAL INJURY AFTER BALLOON ANGIOPLASTY [J].
COY, KM ;
PARK, JC ;
FISHBEIN, MC ;
LAAS, T ;
DIAMOND, GA ;
ADLER, L ;
MAURER, G ;
SIEGEL, RJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (03) :692-700
[4]  
Di Mario C, 1992, J Am Soc Echocardiogr, V5, P135
[5]   CONTRIBUTION OF LOCALIZED CALCIUM DEPOSITS TO DISSECTION AFTER ANGIOPLASTY - AN OBSERVATIONAL STUDY USING INTRAVASCULAR ULTRASOUND [J].
FITZGERALD, PJ ;
PORTS, TA ;
YOCK, PG .
CIRCULATION, 1992, 86 (01) :64-70
[6]   CLASSIFICATION OF MORPHOLOGIC EFFECTS OF PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY ASSESSED BY INTRAVASCULAR ULTRASOUND [J].
GERBER, TC ;
ERBEL, R ;
GORGE, G ;
GE, J ;
RUPPRECHT, HJ ;
MEYER, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (20) :1546-1554
[7]   ARTERIAL-WALL CHARACTERISTICS DETERMINED BY INTRAVASCULAR ULTRASOUND IMAGING - AN INVITRO STUDY [J].
GUSSENHOVEN, EJ ;
ESSED, CE ;
LANCEE, CT ;
MASTIK, F ;
FRIETMAN, P ;
VANEGMOND, FC ;
REIBER, J ;
BOSCH, H ;
VANURK, H ;
ROELANDT, J ;
BOM, N .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (04) :947-952
[8]  
GUSSENHOVEN EJ, 1993, INTRAVASCULAR ULTRAS, P157
[9]  
Gussenhoven W J, 1989, Int J Card Imaging, V4, P105, DOI 10.1007/BF01745140
[10]   MORPHOLOGICAL EFFECTS OF CORONARY BALLOON ANGIOPLASTY INVIVO ASSESSED BY INTRAVASCULAR ULTRASOUND IMAGING [J].
HONYE, J ;
MAHON, DJ ;
JAIN, A ;
WHITE, CJ ;
RAMEE, SR ;
WALLIS, JB ;
ALZARKA, A ;
TOBIS, JM .
CIRCULATION, 1992, 85 (03) :1012-1025