Improvement in antithrombogenicity in a centrifugal pump with self wash-out structure for long-term use

被引:10
作者
WAKISAKA, Y
TAENAKA, Y
TATSUMI, E
ARAKI, K
MASUZAWA, T
NAKATANI, T
BABA, Y
ANAI, H
EYA, K
TODA, K
TAKANO, H
机构
关键词
D O I
10.1097/00002480-199507000-00029
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Antithrombogenicity in the initial (NCVC-1) centrifugal pump (CP) developed at our institute is provided by a central balancing hole in an impeller that promotes self wash-out blood flow (Sf) around the impeller. The current CP (NCVC-2) was ameliorated to obtain better antithrombogenicity, where the balancing hole diameter was widened to improve Sf velocity (Vsf), and the edge of a thrust bearing (TB) was rounded to minimize flow separation. Effects of the modifications were assessed in in vitro and in vivo studies, The Vsf of the NCVC-1 and the NCVC-2 evaluated by Doppler velocimeter were 12.8 and 22.1 cm/sec, respectively, Flow around the TB visualized by a light cutting method confirmed less flow stagnation in the NCVC-2. In vivo anti-thrombogenicity of the CPs was investigated in three goats. A pulsatile VAD (P-VAD) was installed paracorporeally between the left atrium and the aorta. After 3 weeks, the P-VAD was exchanged for the NCVC-2 and the NCVC-1 in sequence. Each CP was driven for 1 week and disassembled. No anticoagulation was used, except for heparin injection at pump exchange. Thrombus at the TB was found in the two NCVC-1s, and two little thrombi were on the impeller of another NCVC-1, whereas a thrombus smaller than 1 mm(3) at the TB was noted in one NCVC-2, These results indicate that the NCVC-2 has better antithrombogenicity and Sf around the impeller and the TB.
引用
收藏
页码:M350 / M355
页数:6
相关论文
empty
未找到相关数据