CT DATA-ACQUISITION AS A BASIS FOR MODERN DIAGNOSIS AND THERAPY IN MAXILLOFACIAL SURGERY

被引:12
作者
FALK, A
GIELEN, S
HEUSER, L
机构
[1] Institute of Radiology and Nuclear Medicine, University Hospital Knappschaftskrankenhaus, Bochum
关键词
CONVENTIONAL COMPUTED TOMOGRAPHY; SPIRAL COMPUTED TOMOGRAPHY; INCREMENT; ARTIFACTS; MAXILLOFACIAL SURGERY;
D O I
10.1016/S0901-5027(05)80865-3
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
More than 20 years after its introduction computed tomography has become a routine method in the diagnostic work-up of traumatic and neoplastic disorders in maxillofacial surgery. Short acquisition times of 1 sec per image reduce motion artifacts and permit increased numbers of patient examinations. Different reconstruction algorithms allow specific evaluation of either osseous or soft tissue structures. By adjusting window center and window width the 2048 attenuation values can be transformed into the visible gray scale. Single sequential slices - as employed in routine diagnosis of tumor patients - still yield the highest image quality. Fast volume data acquisition associated with spiral scanning proves to be of special benefit in trauma patients. The resulting data set can be used for reconstruction of single slices, 2-dimensional (2D) reformated images in all planes and 3D surface rendering.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 17 条
[1]  
BARNES GT, 1989, COMPUTED BODY TOMOGR, P1
[3]   COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[4]   COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022
[5]  
HUPKE R, 1992, ADV CT, V2, P211
[6]  
KALENDER W, 1990, ADV CT, V1, P55
[7]  
KALENDER W, 1992, ADV CT, V2, P87
[8]  
Kalender W A, 1990, Rontgenpraxis, V43, P323
[9]   SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION [J].
KALENDER, WA ;
SEISSLER, W ;
KLOTZ, E ;
VOCK, P .
RADIOLOGY, 1990, 176 (01) :181-183
[10]   PHYSICAL PERFORMANCE-CHARACTERISTICS OF SPIRAL CT SCANNING [J].
KALENDER, WA ;
POLACIN, A .
MEDICAL PHYSICS, 1991, 18 (05) :910-915