SIMULTANEOUS DETERMINATION OF FOLLICLE-STIMULATING-HORMONE AND LUTEINIZING-HORMONE USING A MULTIANALYTE IMMUNOSENSOR

被引:54
作者
PRITCHARD, DJ [1 ]
MORGAN, H [1 ]
COOPER, JM [1 ]
机构
[1] UNIV GLASGOW,DEPT ELECTR & ELECT ENGN,GLASGOW G12 8QQ,LANARK,SCOTLAND
基金
英国生物技术与生命科学研究理事会;
关键词
BIOSENSORS; MULTIANALYTE; PROTEIN PATTERNING; ANTIBODIES; HORMONES;
D O I
10.1016/0003-2670(95)00113-E
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A technique is described for constructing a multianalyte immunosensor for the quantitative determination of the human gonadotrophin hormones follicle stimulating hormone (FSH) and luteinising hormone (LH). The assay is a based upon the electrochemical detection of two horseradish peroxidase labelled antibodies using a ferrocene mediated system. The detection limits (3 sigma) were 2.1 U l(-1) FSH and 1.8 U l(-1) LH. There was negligible cross reactivity. Results obtained with the biosensor showed a good correlation with those obtained using an established clinical diagnostic technique based upon dissociation-enhanced lanthanide fluorometric immunoassay. This is the first report of an electrochemical multianalyte immunosensor for high molecular weight analytes.
引用
收藏
页码:251 / 256
页数:6
相关论文
共 19 条
[1]   FABRICATION OF SURFACES RESISTANT TO PROTEIN ADSORPTION AND APPLICATION TO 2-DIMENSIONAL PROTEIN PATTERNING [J].
BHATIA, SK ;
TEIXEIRA, JL ;
ANDERSON, M ;
SHRIVERLAKE, LC ;
CALVERT, JM ;
GEORGER, JH ;
HICKMAN, JJ ;
DULCEY, CS ;
SCHOEN, PE ;
LIGLER, FS .
ANALYTICAL BIOCHEMISTRY, 1993, 208 (01) :197-205
[2]   MICROPATTERNING PROTEINS AND SYNTHETIC PEPTIDES ON SOLID SUPPORTS - A NOVEL APPLICATION FOR MICROELECTRONICS FABRICATION TECHNOLOGY [J].
BRITLAND, S ;
PEREZARNAUD, E ;
CLARK, P ;
MCGINN, B ;
CONNOLLY, P ;
MOORES, G .
BIOTECHNOLOGY PROGRESS, 1992, 8 (02) :155-160
[3]  
BUECHLER KF, 1992, CLIN CHEM, V38, P1678
[4]   BIOELECTRONIC INTERFACING - MICRO-FABRICATION AND NANOFABRICATION TECHNIQUES FOR GENERATING PREDETERMINED MOLECULAR ARRAYS [J].
CONNOLLY, P .
TRENDS IN BIOTECHNOLOGY, 1994, 12 (04) :123-127
[5]   A METHOD FOR ESTIMATION OF HYDROGEN-PEROXIDE BASED ON MEDIATED ELECTRON-TRANSFER REACTIONS OF PEROXIDASES AT ELECTRODES [J].
FREW, JE ;
HARMER, MA ;
HILL, HAO ;
LIBOR, SI .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1986, 201 (01) :1-10
[6]   ELECTROCHEMICAL BIOSENSORS [J].
FREW, JE ;
HILL, HAO .
ANALYTICAL CHEMISTRY, 1987, 59 (15) :A933-&
[7]  
GOWENLOCK AH, 1988, VARLEYS PRACTICAL BI
[8]  
Green N M, 1975, Adv Protein Chem, V29, P85, DOI 10.1016/S0065-3233(08)60411-8
[9]  
Hall EAH, 1990, BIOSENSORS
[10]   EXTENDING THE DETECTION LIMIT OF SOLID-PHASE ELECTROCHEMICAL ENZYME-IMMUNOASSAY TO THE ATTOMOLE LEVEL [J].
JENKINS, SH ;
HEINEMAN, WR ;
HALSALL, HB .
ANALYTICAL BIOCHEMISTRY, 1988, 168 (02) :292-299