APPLICATION OF INTERLACED ORIENTATION SLICES IN ORTHOGONAL PLANES FOR SYSTEM CHARACTERIZATION AND AS A BASIS FOR STANDARDIZED BRAIN SECTIONS IN MULTIPLE-OBLIQUE POSITIONS

被引:8
作者
BOESCH, C
MARTIN, E
机构
[1] University Children's Hospital, Zürich
关键词
D O I
10.1002/mrm.1910150303
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Interlaced acquisition of orthogonal spin‐echo orientation slices yields saturated intersection lines, which are available for a system characterization (spatial distortions, slice thickness, chemical shift displacement). Additionally, these scout images contain complete information for a transformation of the gradient coordinate system into the oblique coordinate system of the patient, bearing standardized planes for improved image interpretation and quantitative examinations. © 1990 Academic Press, Inc. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:357 / 371
页数:15
相关论文
共 11 条
  • [1] COMBINED APPLICATION OF MR IMAGING AND SPECTROSCOPY IN NEONATES AND CHILDREN - INSTALLATION AND OPERATION OF A 2.35-T SYSTEM IN A CLINICAL SETTING
    BOESCH, C
    MARTIN, E
    [J]. RADIOLOGY, 1988, 168 (02) : 481 - 488
  • [2] BOESCH C, 1988, RADIOLOGY, V169, P384
  • [3] BOESCH C, 1988, 7TH P ANN M SOC MAGN, V2, P699
  • [4] BROSNTEIN IN, 1969, TASCHENBUCH MATH
  • [5] MAGNETIC-RESONANCE IMAGING OF THE HEART - POSITIONING AND GRADIENT ANGLE SELECTION FOR OPTIMAL IMAGING PLANES
    DINSMORE, RE
    WISMER, GL
    LEVINE, RA
    OKADA, RD
    BRADY, TJ
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) : 1135 - 1142
  • [6] OBLIQUE PLANES OF SECTION IN MR IMAGING
    EDELMAN, RR
    STARK, DD
    SAINI, S
    FERRUCCI, JT
    DINSMORE, RE
    LADD, W
    BRADY, TJ
    [J]. RADIOLOGY, 1986, 159 (03) : 807 - 810
  • [7] GATED CARDIAC MAGNETIC-RESONANCE STRUCTURAL IMAGING - OPTIMIZATION BY ELECTRONIC AXIAL ROTATION
    FEIGLIN, DH
    GEORGE, CR
    MACINTYRE, WJ
    ODONNELL, JK
    GO, RT
    PAVLICEK, W
    MEANEY, TF
    [J]. RADIOLOGY, 1985, 154 (01) : 129 - 132
  • [8] KORNN GA, 1961, MATH HDB SCI ENG
  • [9] THE EFFICACY OF TILTED AXIAL MRI OF THE CNS
    RUNGE, VM
    OSBORNE, MA
    WOOD, ML
    WOLPERT, SM
    KWAN, E
    KAUFMAN, DM
    [J]. MAGNETIC RESONANCE IMAGING, 1987, 5 (06) : 421 - 430
  • [10] VARYING GRADIENT ANGLES AND OFFSETS TO OPTIMIZE IMAGING PLANES IN MR
    SLONE, RM
    BUCK, LL
    FITZSIMMONS, JR