BIOCOMPATIBILITY OF VARIOUS INDWELLING DOUBLE-J STENTS

被引:65
作者
CORMIO, L
TALJA, M
KOIVUSALO, A
MAKISALO, H
WOLFF, H
RUUTU, M
机构
[1] UNIV HELSINKI, CENT HOSP, DEPT SURG 4, HELSINKI, FINLAND
[2] PAIJAT HAME CENT HOSP, DEPT SURG, UROL UNIT, LAHTI, FINLAND
[3] UNIV HELSINKI, INST PATHOL, HELSINKI, FINLAND
[4] UNIV BARI, SCH MED, CHAIR UROL R, BARI, ITALY
基金
芬兰科学院;
关键词
URETER; STENTS; MATERIALS TESTING; BIOCOMPATIBLE MATERIALS;
D O I
10.1097/00005392-199502000-00069
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The biocompatibility of 8 double-J stent brands was evaluated both in vivo, by scanning electron microscopy and histological analysis of pig ureters intubated for 6 weeks, and in vitro, by cell culture methods. In vivo findings showed that superficial epithelial destruction was milder in the ureters intubated with hydrogel-coated stents than in those intubated with other stents. Inflammatory reactive changes were milder in the ureters intubated with silicone stents than in those intubated with other stents, while encrustation was more severe on silicone and Grey(TM) stents than the others. In vitro, the 3 stent brands made of ''modified polyurethane'' were found to be cytotoxic while the others were not. Overall, silicone and hydrogel-coated stents seemed to be more biocompatible than the others, and hydrogel-coated stents to be more suitable than silicone stents for long-term ureteral stenting, being less prone to encrustation. Although they are of an experimental nature, these findings may be of relevance to clinical practice.
引用
收藏
页码:494 / 496
页数:3
相关论文
共 19 条
[1]   ROLE OF URETHRAL ISCHEMIA IN THE DEVELOPMENT OF URETHRAL STRICTURES AFTER CARDIOVASCULAR-SURGERY - A PRELIMINARY-REPORT [J].
ABDELHAKIM, A ;
HASSOUNA, M ;
TEIJEIRA, J ;
ELHILALI, M .
JOURNAL OF UROLOGY, 1984, 131 (06) :1077-1079
[2]   TECHNIQUES FOR THE PRESERVATION OF 3-DIMENSIONAL STRUCTURE IN PREPARING SPECIMENS FOR THE ELECTRON MICROSCOPE [J].
ANDERSON, TF .
TRANSACTIONS OF THE NEW YORK ACADEMY OF SCIENCES, 1951, 13 (04) :130-134
[3]   CELLULAR TOXICITY OF URINARY CATHETERS [J].
GRAHAM, DT ;
MARK, GE ;
POMEROY, AR .
MEDICAL JOURNAL OF AUSTRALIA, 1983, 1 (10) :456-459
[4]   THE DEVELOPMENT OF SYNTHETIC-POLYMERS THAT RESIST ENCRUSTATION ON EXPOSURE TO URINE [J].
HOLMES, SAV ;
CHENG, C ;
WHITFIELD, HN .
BRITISH JOURNAL OF UROLOGY, 1992, 69 (06) :651-655
[5]   EVALUATION OF POLYMERIC MATERIALS FOR ENDOUROLOGICAL DEVICES - EMERGING IMPORTANCE OF HYDROGELS [J].
MARDIS, HK .
SEMINARS IN INTERVENTIONAL RADIOLOGY, 1987, 4 (01) :36-45
[6]  
MARDIS HK, 1992, PROBL UROL, V6, P296
[7]   THE EFFECTS OF VARIOUS INDWELLING URETERAL CATHETER MATERIALS ON THE NORMAL CANINE URETER [J].
MARX, M ;
BETTMANN, MA ;
BRIDGE, S ;
BRODSKY, G ;
BOXT, LM ;
RICHIE, JP .
JOURNAL OF UROLOGY, 1988, 139 (01) :180-185
[8]   THE ASSESSMENT OF URINARY CATHETER TOXICITY USING CELL-CULTURES - VALIDATION BY COMPARISON WITH AN ANIMAL-MODEL [J].
NACEY, JN ;
HORSFALL, DJ ;
DELAHUNT, B ;
MARSHALL, VR .
JOURNAL OF UROLOGY, 1986, 136 (03) :706-709
[9]   THE EFFECTS OF DOUBLE-J STENTING ON UNOBSTRUCTED URETERS - AN EXPERIMENTAL AND CLINICAL-STUDY [J].
RAMSAY, JWA ;
PAYNE, SR ;
GOSLING, PT ;
WHITFIELD, HN ;
WICKHAM, JEA ;
LEVISON, DA .
BRITISH JOURNAL OF UROLOGY, 1985, 57 (06) :630-634
[10]   CYTO-TOXICITY OF LATEX URINARY CATHETERS [J].
RUUTU, M ;
ALFTHAN, O ;
TALJA, M ;
ANDERSSON, LC .
BRITISH JOURNAL OF UROLOGY, 1985, 57 (01) :82-87