A portable flow injection analyzer for use with ion-selective electrodes

被引:15
作者
Dimitrakopoulos, T
Alexander, PW
Hibbert, DB
Cherkson, L
Morgan, J
机构
[1] UNIV TASMANIA,DEPT PHYS SCI,LAUNCESTON,TAS 7250,AUSTRALIA
[2] UNIV NEW S WALES,DEPT ANALYT CHEM,KENSINGTON,NSW 2033,AUSTRALIA
关键词
portable instrumentation; battery powered; flow injection; iodide ion-selective electrode;
D O I
10.1002/elan.1140071204
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A battery powered portable monitor based on flow injection potentiometry has been developed weighing 1.8 kg, constructed in a carry-case and applicable for remote site monitoring. Portability has been achieved using a light weight rechargeable Ni-Cd battery-pack (7.2 V). The flow injection manifold incorporates a low powered peristaltic pump, a wall-jet type flow cell containing a commercial iodide ion-selective electrode and a Ag/AgCl reference electrode connected to an analog-to-digital converter, also powered by the battery pack, and a RS232-serial output to a notebook computer for real-time data display and storage. The system performance has been evaluated using the iodide ion-selective electrode, including the effects of various carrier solutions on sample peak heights, peak widths and the calibration slopes and working ranges. Fast response with peak widths as low as approximately 10 s. was observed with near Nernstian response of -53.8 mV change per activity decade and a log-linear range between 5 x 10(-6) and 1 x 10(-2) M in a 0.1 M sodium acetate carrier stream.
引用
收藏
页码:1118 / 1120
页数:3
相关论文
共 15 条
[1]   RAPID-FLOW CONTINUOUS ANALYSIS WITH ION-SELECTIVE ELECTRODES [J].
ALEXANDER, PW ;
SEEGOPAUL, P .
ANALYTICAL CHEMISTRY, 1980, 52 (14) :2403-2406
[2]   RAPID FLOW-ANALYSIS WITH INDUCTIVELY COUPLED PLASMA ATOMIC-EMISSION SPECTROSCOPY USING A MICRO-INJECTION TECHNIQUE [J].
ALEXANDER, PW ;
FINLAYSON, RJ ;
SMYTHE, LE ;
THALIB, A .
ANALYST, 1982, 107 (1280) :1335-1342
[3]   AUTOMATED PROTEIN DETERMINATION WITH ION-SELECTIVE MEMBRANE ELECTRODES [J].
ALEXANDER, PW ;
RECHNITZ, GA .
ANALYTICAL CHEMISTRY, 1974, 46 (07) :860-865
[4]  
Andrew Kevin N., 1994, Analytical Chemistry, V66, p917A
[6]   HIGH-SPEED POTENTIOMETRIC ANALYZER EQUIPPED WITH AN ION-SELECTIVE ELECTRODE DETECTOR [J].
HARA, H ;
ISHIO, N ;
TAKAHASHI, K .
ANALYTICA CHIMICA ACTA, 1993, 281 (01) :45-51
[7]   ADSORPTION PHENOMENA OF SILVER-IODIDE BASED ION-SELECTIVE ELECTRODES [J].
HARSANYI, EG ;
TOTH, K ;
POLOS, L ;
PUNGOR, E .
ANALYTICAL CHEMISTRY, 1982, 54 (07) :1094-1097
[8]   CONTINUOUS DETERMINATION OF NITRATE CONCENTRATIONS INSITU [J].
JOHNSON, KS ;
SAKAMOTOARNOLD, CM ;
BEEHLER, CL .
DEEP-SEA RESEARCH PART A-OCEANOGRAPHIC RESEARCH PAPERS, 1989, 36 (09) :1407-1413
[9]   A SUBMERSIBLE FLOW-ANALYSIS SYSTEM [J].
JOHNSON, KS ;
BEEHLER, CL ;
SAKAMOTOARNOLD, CM .
ANALYTICA CHIMICA ACTA, 1986, 179 :245-257
[10]   THEORY AND APPLICATIONS OF ION-SELECTIVE ELECTRODES .7. [J].
KORYTA, J .
ANALYTICA CHIMICA ACTA, 1988, 206 (1-2) :1-48