FLUCTUATIONS IN SYRINGE-PUMP INFUSIONS - ASSOCIATION WITH BLOOD-PRESSURE VARIATIONS IN INFANTS

被引:39
作者
CAPES, DF
DUNSTER, KR
SUNDERLAND, VB
MCMILLAN, D
COLDITZ, PB
MCDONALD, C
机构
[1] CURTIN UNIV TECHNOL,SCH PHARM,BENTLEY,WA,AUSTRALIA
[2] ROYAL HOSP WOMEN,PERINATAL RES CTR,BRISBANE,QLD,AUSTRALIA
[3] CURTIN UNIV TECHNOL,SCH PHARM,BENTLEY,WA,AUSTRALIA
关键词
DEVICES; DOPAMINE; DRUG ADMINISTRATION RATE; FLOW; HYPOTENSION; INJECTIONS; PEDIATRICS; SYMPATHOMIMETIC AGENTS; SYRINGES;
D O I
10.1093/ajhp/52.15.1646
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Flow continuity of two brands of syringe pumps and four brands of syringes was studied as a possible cause of hemodynamic fluctuations observed in neonates. Cyclical fluctuations were observed in the blood pressure of 14 neonates receiving dopamine infusions by syringe pump at flow rates from 0.2 to 1 mL/hr. Atom 235 and IVAC 770 pumps and various sizes of Terumo, Becton Dickinson, Omnifix, and IVAC syringes were evaluated. Flow continuity was assessed by using a gravimetric technique. The force needed to initiate and maintain syringe plunger motion was also measured. Noncontinuous flow was encountered most commonly with Temmo syringes, which delivered boluses at regular intervals at flow rates up to 5 ml/hr. The interval was dependent on flow rate and was similar to the time between the blood pressure fluctuations observed clinically. The syringe plunger force exhibited regular fluctuations indicative of the plunger sticking, and simultaneous measurement of flow established a direct temporal relationship with boluses. The other syringes tested did not exhibit such fluctuations. No differences were found between the two syringe pumps. Syringe plunger sticking, resulting in intermittent boluses and potential blood pressure fluctuations, may occur at low flow rates and with certain syringe brands. This appeared to be the cause of hemodynamic fluctuations in neonates receiving dopamine infusions.
引用
收藏
页码:1646 / 1653
页数:8
相关论文
共 23 条
[1]  
BHATTMEHTA V, 1989, PHARMACOTHERAPY, V9, P303
[2]  
CAPES DF, IN PRESS J PARENTER
[3]  
CARLISLE A, 1992, SEP AUSTR PHYS SCI E
[4]   CYCLICAL VARIATION OF BLOOD-PRESSURE AND HEART-RATE IN NEONATES [J].
CUNNINGHAM, S ;
DEERE, S ;
MCINTOSH, N .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1993, 69 (01) :64-67
[5]   EFFECT OF INFUSION PUMP FILL-STROKE FLOW INTERRUPTIONS ON SIMULATED SERUM CONCENTRATIONS OF SHORT HALF-LIFE DRUGS [J].
FLOYD, RA .
AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1984, 41 (11) :2399-2400
[6]   CLINICAL EVENTS IN ASSOCIATION WITH TIMING OF INTRAVENTRICULAR HEMORRHAGE IN PRETERM INFANTS [J].
FUNATO, M ;
TAMAI, H ;
NOMA, K ;
KURITA, T ;
KAJIMOTO, Y ;
YOSHIOKA, Y ;
SHIMADA, S .
JOURNAL OF PEDIATRICS, 1992, 121 (04) :614-619
[7]   INFLUENCE OF INFUSION PUMPS ON THE PHARMACOLOGICAL RESPONSE TO NITROPRUSSIDE [J].
HURLBUT, JC ;
THOMPSON, S ;
REED, MD ;
BLUMER, JL ;
ERENBERG, A ;
LEFF, RD .
CRITICAL CARE MEDICINE, 1991, 19 (01) :98-101
[8]   INFLUENCE OF INFUSION-PUMP OPERATION AND FLOW-RATE ON HEMODYNAMIC STABILITY DURING EPINEPHRINE INFUSION [J].
KLEM, SA ;
FARRINGTON, JM ;
LEFF, RD .
CRITICAL CARE MEDICINE, 1993, 21 (08) :1213-1217
[9]  
MANN HJ, 1988, CLIN PHARMACY, V7, P214
[10]   PHYSIOLOGIC OSCILLATIONS PRODUCED BY AN INFUSION PUMP [J].
SCHULZE, KF ;
GRAFF, M ;
SCHIMMEL, MS ;
SCHENKMAN, A ;
ROHAN, P .
JOURNAL OF PEDIATRICS, 1983, 103 (05) :796-798