A FLOW VISUALIZATION STUDY OF THE NCVC CENTRIFUGAL BLOOD PUMP

被引:10
作者
ARAKI, K
TAENAKA, Y
MASUZAWA, T
TATSUMI, E
WAKISAKA, Y
WATARI, M
NAKATANI, T
AKAGI, H
BABA, Y
ANAI, H
PARK, YH
EYA, K
TODA, K
TAKANO, H
机构
[1] Department of Artificial Organs, National Cardiovascular Center Research Institute, Osaka, suita City
关键词
CENTRIFUGAL PUMP; FLOW VISUALIZATION; TRACER METHOD; OIL FILM METHOD; INJECTION STREAK LINE METHOD;
D O I
10.1111/j.1525-1594.1994.tb03397.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A compact centrifugal pump, NCVC-1, has an open-type impeller with 6 curved vanes, and it is characterized by no shaft and no seal. A tunnel is placed in the center of the impeller-rotor assembly to irrigate the back space behind the rotor. To evaluate the flow, we performed 3 visualization methods: tracer, oil film, and injection streak line method. The flow, observed by the tracer method in NCVC-1, indicated little turbulence along vanes. A volute chamber proved effective to reduce vortex formation in the outlet. Oil film pattern revealed no flow separation on vanes at 5 L/min. Washout flow behind the rotor is essential to prevent thrombus formation and was shown as inward spiral flow without any stagnation. These data suggested that a combination of visualization techniques was useful to analyze various flow conditions, and the NCVC-1 has excellent flow characteristics with little turbulence and little flow stagnation, which must be beneficial to low hemolysis and high antithrombogenicity.
引用
收藏
页码:669 / 672
页数:4
相关论文
共 4 条
[1]  
Affeld K, 1986, ASAIO Trans, V32, P269
[2]  
ARAKI K, 1993, ARTIF ORGANS, V17, P307
[3]   PARTICLE STREAK VELOCITY-FIELD MEASUREMENTS IN A TWO-DIMENSIONAL MIXING LAYER [J].
DIMOTAKIS, PE ;
DEBUSSY, FD ;
KOOCHESFAHANI, MM .
PHYSICS OF FLUIDS, 1981, 24 (06) :995-999
[4]  
Taenaka Y, 1992, ASAIO J, V38, pM577, DOI 10.1097/00002480-199207000-00101