4-FRAME GATED WOLTER X-RAY MICROSCOPE

被引:21
作者
ELLIS, RJ
TREBES, JE
PHILLION, DW
KILKENNY, JD
GLENDINNING, SG
WIEDWALD, JD
LEVESQUE, RA
机构
[1] Lawrence Livermore National Laboratory, Livermore, CA 94550
关键词
D O I
10.1063/1.1141822
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A four-frame gated x-ray microscope has been developed for the study of inertial confinement fusion. Four x-ray images were obtained by introducing flat, grazing incidence mirrors into the converging annular ray bundle of a previously developed Wolter x-ray microscope system. Gated microchannel plate detectors and associated drive electronics were developed which can record a sequence of gated images of less than 100 ps duration, with an interframe timing accuracy of ±50 ps. The resulting diagnostic can record one time-resolved image and four gated images simultaneously. The system is described, and images of an imploding capsule are presented which demonstrate the unique capabilities of the instrument.