CLINICAL-EVALUATION OF A HIGH-RESOLUTION (2.6-MM) POSITRON EMISSION TOMOGRAPH

被引:21
作者
VALK, PE
JAGUST, WJ
DERENZO, SE
HUESMAN, RH
GEYER, AB
BUDINGER, TF
机构
[1] UNIV CALIF DAVIS,DEPT NEUROL,DAVIS,CA 95616
[2] UNIV CALIF BERKELEY LAWRENCE BERKELEY LAB,BERKELEY,CA 94720
关键词
D O I
10.1148/radiology.176.3.2389037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The intrinsic resolution of the Donner 600-crystal positron emission tomograph (PET 600) is 2.6 mm full width at half maximum (FWHM) inplane and 6 mm FWHM axially. More than 100 patients with glioma, radiation necrosis, Alzheimer disease, or epilepsy have been studied with this system. Approximately 1 million events are acquired in 15 minutes, starting 1 hour after injection of 10 mCi (370 MBq) of fluorine-18-fluorodeoxyglucose. Normal structures as small as the superior colliculi and the external capsule have been resolved. Improved separation of the cortical ribbon from adjacent white matter has allowed more accurate determination of cortical metabolic rate. In two of 15 patients undergoing evaluation for recurrent glioma, the PET 600 images showed tumor uptake that was not apparent on a lower-resolution study. A high-activity orbiting transmission source with electronic collimation allows accurate, short-duration transmission measurements to be made after radiopharmaceutical administration. The anatomic detail seen on the transmission images can be used for reproducible patient positioning with an accuracy of 1-2 mm perpendicular to the image plane. These findings demonstrate the practicality and clinical effectiveness of high-resolution positron emission tomography.
引用
收藏
页码:783 / 790
页数:8
相关论文
共 18 条
[1]  
BUDINGER TF, 1984, ANN NEUROL, V15, pS35
[2]  
DERENZO S E, 1989, Journal of Nuclear Medicine, V30, P745
[3]   IMAGING PROPERTIES OF A POSITRON TOMOGRAPH WITH 280-BGO-CRYSTALS [J].
DERENZO, SE ;
BUDINGER, TF ;
HUESMAN, RH ;
CAHOON, JL ;
VULETICH, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) :81-89
[4]   A POSITRON TOMOGRAPH WITH 600-BGO CRYSTALS AND 2.6 MM RESOLUTION [J].
DERENZO, SE ;
HUESMAN, RH ;
CAHOON, JL ;
GEYER, AB ;
MOSES, WW ;
UBER, DC ;
VULETICH, T ;
BUDINGER, TF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :659-664
[5]  
DERENZO SE, 1988, J NUCL MED, V29, P761
[6]  
GRADY CL, 1989, J NUCL MED, V30, P1386
[7]   BRACHYTHERAPY OF RECURRENT MALIGNANT BRAIN-TUMORS WITH REMOVABLE HIGH-ACTIVITY I125 SOURCES [J].
GUTIN, PH ;
PHILLIPS, TL ;
WARA, WM ;
LEIBEL, SA ;
HOSOBUCHI, Y ;
LEVIN, VA ;
WEAVER, KA ;
LAMB, S .
JOURNAL OF NEUROSURGERY, 1984, 60 (01) :61-68
[8]   REGIONAL KINETIC CONSTANTS AND CEREBRAL METABOLIC-RATE FOR GLUCOSE IN NORMAL HUMAN VOLUNTEERS DETERMINED BY DYNAMIC POSITRON EMISSION TOMOGRAPHY OF [F-18]-2-FLUORO-2-DEOXY-D-GLUCOSE [J].
HEISS, WD ;
PAWLIK, G ;
HERHOLZ, K ;
WAGNER, R ;
GOLDNER, H ;
WIENHARD, K .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1984, 4 (02) :212-223
[9]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .1. EFFECT OF OBJECT SIZE [J].
HOFFMAN, EJ ;
HUANG, SC ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (03) :299-308
[10]   A NEW FAST ALGORITHM FOR THE EVALUATION OF REGIONS OF INTEREST AND STATISTICAL UNCERTAINTY IN COMPUTED-TOMOGRAPHY [J].
HUESMAN, RH .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (05) :543-552