CALCULATION OF WATER AIR STOPPING-POWER RATIOS USING EGS4 WITH EXPLICIT TREATMENT OF ELECTRON-POSITRON DIFFERENCES

被引:31
作者
MALAMUT, C [1 ]
ROGERS, DWO [1 ]
BIELAJEW, AF [1 ]
机构
[1] NATL RES COUNCIL CANADA,INST NATL MEASUREMENT STAND,OTTAWA K1A 0R6,ONTARIO,CANADA
关键词
D O I
10.1118/1.596594
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Using the EGS4 Monte Carlo simulation program, a general purpose code has been written to calculate Bragg-Gray and Spencer-Attix stopping-power ratios for use in radiation dosimetry. The stopping-power ratios can be calculated in any material in any region in a general cylindrical geometry with a large number of source geometries possible. The calculations take into account for the first time the differences between the stopping powers and the inelastic scattering of positrons and electrons. The results show that previous calculations ignoring these effects were accurate. The present results agree, typically within 0.1%, with the Spencer-Attix water-to-air stopping-power ratios for broad parallel beams of electrons given in the AAPM and IAEA protocols except at the surface where the present calculations follow the buildup of secondary electrons in more detail and see a 2% reduction in the stopping-power ratios.
引用
收藏
页码:1222 / 1228
页数:7
相关论文
共 25 条
[1]   DEPTH DOSE AND STOPPING-POWER DATA FOR MONOENERGETIC ELECTRON-BEAMS [J].
ANDREO, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1990, 51 (02) :107-121
[2]   STOPPING-POWER RATIOS AND THEIR UNCERTAINTIES FOR CLINICAL ELECTRON-BEAM DOSIMETRY [J].
ANDREO, P ;
FRANSSON, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (12) :1847-1861
[3]   STOPPING POWER DATA FOR HIGH-ENERGY PHOTON BEAMS [J].
ANDREO, P ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (08) :839-858
[4]  
[Anonymous], 1983, MED PHYS, V10, P741
[5]  
[Anonymous], [No title captured]
[6]  
BERGER MJ, 1975, BIOMEDICAL DOSIMETRY, P589
[7]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[8]  
CHILTON AB, 1978, HEALTH PHYS, V34, P715
[9]  
CHILTON AB, 1979, HLTH PHYS, V35, P637
[10]   ON THE SELECTION OF STOPPING-POWER AND MASS ENERGY-ABSORPTION COEFFICIENT RATIOS FOR HIGH-ENERGY X-RAY DOSIMETRY [J].
CUNNINGHAM, JR ;
SCHULZ, RJ .
MEDICAL PHYSICS, 1984, 11 (05) :618-623