VOLUME SCANNING IN THE EVALUATION OF FETAL MALFORMATIONS - A NEW DIMENSION IN PRENATAL-DIAGNOSIS

被引:202
作者
MERZ, E
BAHLMANN, F
WEBER, G
机构
[1] Department of Obstetrics and Gynecology, Center for Ultrasound Diagnosis and Prenatal Therapy, Johannes Gutenberg University, Mainz
关键词
3-DIMENSIONAL ULTRASOUND; PRENATAL DIAGNOSIS; FETAL MALFORMATIONS;
D O I
10.1046/j.1469-0705.1995.05040222.x
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Three-dimensional ultrasound examination was performed in 204 patients with a fetal malformation detected by conventional ultrasound. The patients were examined between 13 and 40 weeks of gestation. The ultrasound equipment used was a Combison 330 and a Combison 530 (Kretztechnik, Austria) with an abdominal Voluson sector transducer (3.5/5 MHz) (Kretztechnik, Austria). This ultrasound system can provide a high-quality three-dimensional surface or translucency image of fetal structures similar to that of a photograph or an X-ray image within seconds without an additional expensive work-station. Of the 204 patients examined with three-dimensional ultrasound, this technique proved advantageous in demonstrating fetal defects in 62% (127/204). In 36% (73/204), the three-dimensional technique gave the same information and in four fetuses with a cardiac malformation (2%), the three-dimensional technique was disadvantageous, due to movement artefacts during data acquisition. The technical advantages and problems of this three-dimensional technique are demonstrated.
引用
收藏
页码:222 / 227
页数:6