STRATEGY FOR TRACE-METAL DETERMINATION IN SEAWATER BY ANODIC-STRIPPING VOLTAMMETRY USING A COMPUTERIZED MULTI-TIME-DOMAIN MEASUREMENT METHOD

被引:15
作者
BOND, AM
HERITAGE, ID
THORMANN, W
机构
[1] Deakin Univ, Waurn Ponds, Aust, Deakin Univ, Waurn Ponds, Aust
关键词
ELECTROCHEMISTRY - METALS AND ALLOYS - Measurements;
D O I
10.1021/ac00297a018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A microcomputer-based electrochemistry system has been developed with capabilities of assigning the reliability of different methods of trace analytical voltammetric determinations employing transient techniques such as differential pulse and square wave anodic stripping voltammetry in seawater. The system incorporates peak height, width, and potential measurements as a function of time and concentration to construct a comprehensive data base. Seawater samples are compared against this reference to detect if interference from surfactants or other sources are present and to decide (1) whether the direct calibration method is appropriate, (II) whether the more time-consuming standard addition method is satisfactory, or (iii) if UV irradiation is required to eliminate matrix effects.
引用
收藏
页码:1063 / 1066
页数:4
相关论文
共 23 条
[1]   INFLUENCE OF LABORATORY ENVIRONMENT ON THE PRECISION AND ACCURACY OF TRACE-ELEMENT ANALYSIS [J].
ADELOJU, SB ;
BOND, AM .
ANALYTICAL CHEMISTRY, 1985, 57 (08) :1728-1733
[2]   STRIPPING VOLTAMMETRIC DETERMINATION OF SELENIUM IN BIOLOGICAL-MATERIALS BY DIRECT CALIBRATION [J].
ADELOJU, SB ;
BOND, AM ;
BRIGGS, MH ;
HUGHES, HC .
ANALYTICAL CHEMISTRY, 1983, 55 (13) :2076-2082
[3]   STUDIES OF MEDIUM EFFECTS ON THE DPASV DETERMINATION OF CU IN NATURAL-WATERS USING CONVENTIONAL AND FLOW-THROUGH ELECTRODES [J].
ALZAND, IK ;
LANGFORD, CH .
CANADIAN JOURNAL OF CHEMISTRY-REVUE CANADIENNE DE CHIMIE, 1985, 63 (03) :643-650
[4]   MICROPROCESSOR-CONTROLLED INSTRUMENT FOR THE SIMULTANEOUS GENERATION OF SQUARE-WAVE, ALTERNATING-CURRENT, DIRECT-CURRENT, AND PULSE POLAROGRAMS [J].
ANDERSON, JE ;
BOND, AM .
ANALYTICAL CHEMISTRY, 1983, 55 (12) :1934-1939
[5]  
BATLEY G E, 1976, Marine Chemistry, V4, P347, DOI 10.1016/0304-4203(76)90020-7
[6]   EFFECT OF DISSOLVED ORGANICS ON STRIPPING VOLTAMMETRY OF SEAWATER [J].
BATLEY, GE ;
FLORENCE, TM .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1976, 72 (01) :121-126
[7]   RELIABLE USE OF CALIBRATION CURVES IN VOLTAMMETRIC ANALYSIS WITH A NEW TECHNIQUE OF MICROCOMPUTER-BASED DATA EVALUATION [J].
BOND, AM ;
HERITAGE, ID .
ANALYTICAL CHEMISTRY, 1985, 57 (01) :174-179
[8]   DEVELOPMENT OF A MICROPROCESSOR-BASED ELECTROCHEMICAL INSTRUMENT INTERFACED TO A MICROCOMPUTER SYSTEM FOR DIFFERENTIAL-PULSE STRIPPING VOLTAMMETRY IN DIFFERENT TIME DOMAINS [J].
BOND, AM ;
GREENHILL, HB ;
HERITAGE, ID ;
REUST, JB .
ANALYTICA CHIMICA ACTA, 1984, 165 (NOV) :209-216
[9]  
FABRIS JG, 1983, 24 MIN CONS MAR SCI
[10]   CYBERNETIC CONTROL OF AN ELECTROCHEMICAL REPERTOIRE [J].
HE, P ;
AVERY, JP ;
FAULKNER, LR .
ANALYTICAL CHEMISTRY, 1982, 54 (12) :A313-+