DEVELOPMENT OF MOCK CIRCULATION MODELS FOR THE ASSESSMENT OF COUNTERPULSATION SYSTEMS

被引:16
作者
BOWLES, CT
NISHIMURA, K
SHAH, SS
CUMMING, DVE
PEPPER, JR
YACOUB, MH
CLARK, C
PATTISON, CW
机构
[1] NATL HEART & LUNG INST,DEPT CARDIOTHORAC SURG,LONDON SW3,ENGLAND
[2] BRUNEL UNIV,DEPT MFG & ENGN SYST,UXBRIDGE UB8 3PH,MIDDX,ENGLAND
[3] MIDDLESEX HOSP,LONDON W1,ENGLAND
关键词
COUNTERPULSATION; MOCK CIRCULATION; SKELETAL MUSCLE; CIRCULATORY SUPPORT;
D O I
10.1093/cvr/25.11.901
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Study objective -This study entailed the development of mock circulation models to assess and compare the haemodynamic efficacy of extra-aortic counterpulsation (using trained skeletal muscle wrapped around the proximal descending aorta) and conventional intra-aortic balloon counterpulsation. Design and experimental materials - Hydraulic Windkessel type lumped parameter models were used either in conjunction with native skeletal muscle or as a dynamic simulation of counterpulsation. The haemodynamic performance of the wrapped latissimus dorsi muscle of the normal sheep was assessed using an artificial load to simulate the pressurised proximal descending aorta. Mock circulation models of counterpulsation comprised Windkessel compliance chambers, laminar flow resistors, a blood analogue, a prosthetic blood pump, and a purpose made hydraulic counterpulsator. Measurements and main results - An electrically stimulated muscle wrap, 5 cm in length, previously trained for 2 weeks at 3 V and 35 Hz, was assessed for haemodynamic performance in a mock circulation: volume of fluid displaced = 14.1(SD 1.8) ml; pressure increase from 100 mm Hg = 14.9(2.1) mm Hg; external work per contraction cycle = 180(70) mJ; external mean power = 800(100) mW. In a simulation of intra-aortic balloon counterpulsation, haemodynamic benefit (ie, an increase in proximal flow rate and endocardial viability ratio and a reduction in left ventricular stroke power) was assessed with respect to defined parameters. Conclusions - This paper demonstrates the potential of the mock circulation models both for the investigation of muscle wrap performance and for the comparison of extra-aortic muscle with intra. aortic balloon counterpulsation.
引用
收藏
页码:901 / 908
页数:8
相关论文
共 42 条
[1]   SKELETAL-MUSCLE AS THE POTENTIAL POWER SOURCE FOR A CARDIOVASCULAR PUMP - ASSESSMENT INVIVO [J].
ACKER, MA ;
HAMMOND, RL ;
MANNION, JD ;
SALMONS, S ;
STEPHENSON, LW .
SCIENCE, 1987, 236 (4799) :324-327
[2]  
Amsterdam E A, 1981, Cardiovasc Clin, V11, P79
[3]  
Aperia A., 1940, SKAND ARCH PHYSL S, V83, P1
[4]   CARDIAC ASSISTANCE WITH ELECTRICALLY STIMULATED SKELETAL-MUSCLE [J].
BADYLAK, SF ;
STEVENS, L ;
JANAS, W ;
GRAY, MH ;
GEDDES, LA ;
VOORHEES, WD .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1989, 27 (02) :159-162
[6]  
BRISTER SJ, 1985, CAN J SURG, V4, P341
[7]   EXPERIMENTAL SUBENDOCARDIAL ISCHEMIA IN DOGS WITH NORMAL CORONARY-ARTERIES [J].
BUCKBERG, GD ;
ARCHIE, JP ;
FIXLER, DE ;
HOFFMAN, JIE .
CIRCULATION RESEARCH, 1972, 30 (01) :67-+
[8]  
CARPENTIER A, 1985, LANCET, V1, P1267
[9]  
CHACHQUES JC, 1988, CIRCULATION, V78, P203
[10]   DYNAMIC AORTOMYOPLASTY TO ASSIST LEFT-VENTRICULAR FAILURE [J].
CHACHQUES, JC ;
GRANDJEAN, PA ;
FISCHER, EIC ;
LATREMOUILLE, C ;
JEBARA, VA ;
BOURGEOIS, I ;
CARPENTIER, A .
ANNALS OF THORACIC SURGERY, 1990, 49 (02) :225-230