PHOSPHATE-SENSITIVE ENZYME ELECTRODE - A POTENTIAL SENSOR FOR ENVIRONMENT CONTROL

被引:55
作者
DURSO, EM [1 ]
COULET, PR [1 ]
机构
[1] UNIV LYON 1,GENIE ENZYMAT LAB,CNRS UMR 106,43 BD 11 NOVEMBRE 1918,F-69622 VILLEURBANNE,FRANCE
关键词
Enzyme electrode; Phosphate; Waters;
D O I
10.1016/S0003-2670(00)83827-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive enzyme electrode was designed for the assay of phosphate ions. For this purpose, a bienzyme membrane with co-immobilized nucleoside phosphorylase and xanthine oxidase was used with a platinum amperometric electrode for the detection of enzymatically generated hydrogen peroxide. A detection limit of 10-7 M was obtained and phosphate assays could be easily performed in the range 0.1-10 μM, which is of interest in the control of water pollution. © 1990.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 16 条
[1]  
ASSOLANTVINET CH, 1986, ANAL LETT, V19, P875
[2]  
Barman T.E., 1969, ENZYME HDB, VI
[3]  
Bray R. C., 1963, ENZYMES, V7, P533
[4]  
BURGIS MJ, 1987, NATURAL HIST LAKES, P35
[5]  
Engasser J.-M., 1976, APPL BIOCHEM BIOENG, V1, P127
[6]  
Fiske CH, 1925, J BIOL CHEM, V66, P375
[7]  
GOLDSTEIN L, 1968, Z ANAL CHEM FRESENIU, V243, P375
[8]   PHOSPHATE-SELECTIVE ELECTRODE BASED ON IMMOBILIZED ALKALINE-PHOSPHATASE AND GLUCOSE OXIDASE [J].
GUILBAULT, GG ;
NANJO, M .
ANALYTICA CHIMICA ACTA, 1975, 78 (01) :69-80
[9]   DETERMINATION OF INORGANIC-PHOSPHATE BY FLOW-INJECTION METHOD WITH IMMOBILIZED ENZYMES AND CHEMI-LUMINESCENCE DETECTION [J].
KAWASAKI, H ;
SATO, K ;
OGAWA, J ;
HASEGAWA, Y ;
YUKI, H .
ANALYTICAL BIOCHEMISTRY, 1989, 182 (02) :366-370
[10]   HYBRID ELECTRODE FOR THE DETERMINATION OF PHOSPHATES AND POLYPHOSPHATES - APPLICATION TO ENVIRONMENTAL-PROBLEMS [J].
LINDERS, CR ;
VINCKE, BJ ;
PATRIARCHE, GJ .
ANALYTICAL LETTERS PART B-CLINICAL AND BIOCHEMICAL ANALYSIS, 1985, 18 (17) :2195-2208