摘要
Although functional replacement of the heart and lungs by a pump and oxygenator is a widespread surgical procedure, no widely accepted technique for describing gas exchange in oxygenators exists. In this study, 8 types of commercially available membrane oxygenator (2 flat sheet membrane, 4 gas in hollow fibre membrane and 2 blood in hollow fibre membrane) have been studied during clinical cardiopulmonary bypass. O2 transfer increased with blood flow rate but the O2 transfer at a given blood flow was lower than that obtained by the manufacturers in laboratory studies. Overall O2 transfer coefficients were calculated from the ratio of O2, transfer rate to an O2 difference expressed either as an O2 partial pressure or an O2 concentration. Specific O2 transfer coefficients (overall coefficient divided by membrane area) were similar for oxygenators with a flat sheet or gas in hollow fibre membrane configuration. The two types of oxygenator with blood in hollow fibre membranes had significantly lower (P<0.01) specific O2 transfer coefficients. This study shows that oxygenator gas transfer characteristics can be studied in the clinical environment and that O2 transfer coefficients can be related to oxygenator design features.
引用
收藏
页码:167 / 177
页数:11
相关论文
共 23 条
[1]  
ALDER HL, 1975, INTRO PROBABILITY ST
[2]  
CLAYTON RH, 1989, THESIS U NEWCASTLE T
[3]  
COLE PV, 1969, BIOMED ENG, V2, P56
[4]  
CURTIS RM, 1974, T AM SOC ART INT ORG, VA 20, P210
[5]   OXYGEN AND CO2 TRANSFER OF A POLYPROPYLENE DIMPLED MEMBRANE LUNG WITH VARIABLE SECONDARY FLOWS [J].
DORRINGTON, KL ;
RALPH, ME ;
BELLHOUSE, BJ ;
GARDAZ, JP ;
SYKES, MK .
JOURNAL OF BIOMEDICAL ENGINEERING, 1985, 7 (02) :89-99
[6]  
DORRINGTON KL, 1989, OXFORD MED ENG SERIE, V9
[7]  
EBERHART RC, 1978, ARTIF ORGANS, V2, P19
[8]  
GALLETTI PM, 1972, T AM SOC ART INT ORG, V18, P359
[9]   MEMBRANE OXYGENATORS - CURRENT DEVELOPMENTS IN DESIGN AND APPLICATION [J].
GAYLOR, JDS .
JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (06) :541-547
[10]  
GAYLOR JDS, 1976, PHYSL CLIN ASPECTS O