COMPARISON OF MEASURED AND CALCULATED DOSE-RATES IN WATER NEAR I-125 AND IR-192 SEEDS

被引:174
作者
WILLIAMSON, JF
机构
[1] Radiation Oncology Center, Mallinckrodt Institute of Radiology, Washington University Schoot of Medicine, St. Louis, Missouri
关键词
D O I
10.1118/1.596631
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Recent theoretical and experimental work indicates that currently accepted I-125 dosimetry data may overestimate dose in water at 1 cm by 10%-24%. Among the most comprehensive measurements are those of the NCI-sponsored brachytherapy contract participants. Absolute dose rates in water calculated by the Monte Carlo method have been compared with the NCI dose measurements about I-125 and Ir-192 seeds embedded in solid-water phantoms. The photon transport code allows realistic geometric simulation of the complex internal seed structure, the National Institute of Standards and Technology air-kerma strength standardization geometry, and the dose measurement setup. When the appropriate measurement medium and geometry are assumed, agreement between theory and measurement is excellent, within 3% at 1 cm and averaging 3% at larger distances. However, the data do not support the water equivalence of solid water at I-125 energies indicating that solid-water measurements underestimate I-125 specific dose-rate constants in water by 4.3%. Because of its higher ratio of absorption to scatter, I-125 dose distributions measured in solid water are less penetrating (by 35% at 10 cm) than those measured in liquid water. For model 6711, model 6702, and steel-clad Ir-192 seeds, Monte Carlo calculations yielded specific dose-rate constants (assuming liquid water medium) of 0.877, 0.932, and 1.122 cGy cm2 h-1 per unit air-kerma strength, respectively. For I-125, currently accepted values are 18% and 11% larger for the two seed models.
引用
收藏
页码:776 / 786
页数:11
相关论文
共 44 条
[1]  
Anderson LL, 1990, INTERSTITIAL BRACHYT
[2]  
BERGER MJ, 1964, J NUCL MED S1, V15
[3]   TWO-DIMENSIONAL DOSE DISTRIBUTION AROUND A COMMERCIAL I125 SEED [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1988, 15 (01) :56-60
[4]   MONTE-CARLO SIMULATION OF THE DOSE DISTRIBUTION AROUND I-125 SEEDS [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1987, 14 (03) :420-424
[5]   DOSE-RATE DETERMINATION FOR I-125 SEEDS [J].
CHIUTSAO, ST ;
ANDERSON, LL ;
OBRIEN, K ;
SANNA, R .
MEDICAL PHYSICS, 1990, 17 (05) :815-825
[6]  
DILLMAN LT, 1975, NMMIRD10 SOC NUCL ME
[7]  
GLASGOW G P, 1979, Medical Physics (Woodbury), V6, P49, DOI 10.1118/1.594551
[8]  
HILARIS BS, 1975, DHEW FDA768022
[9]  
Hubbell J. H., 1975, Journal of Physical and Chemical Reference Data, V4, P471, DOI 10.1063/1.555523
[10]  
Hubbell J.H, 1969, NSRDSNBS29