BAYESIAN IMAGE ESTIMATION OF DIGITAL CHEST RADIOGRAPHY - INTERDEPENDENCE OF NOISE, RESOLUTION, AND SCATTER FRACTION

被引:16
作者
BAYDUSH, AH [1 ]
FLOYD, CE [1 ]
机构
[1] DUKE UNIV,DEPT BIOMED ENGN,DURHAM,NC 27710
关键词
SCATTER REDUCTION; DIGITAL THORACIC IMAGING; STATISTICAL IMAGE PROCESSING;
D O I
10.1118/1.597563
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Previously, it has been shown that Bayesian image estimation (BIE) can reduce the effects of scattered radiation and improve contrast-to-noise ratios (CNR) in digital radiographs of anthropomorphic chest phantoms by improving contrast while constraining noise. Here, the use of BIE as a noise reduction technique is reported. An anthropomorphic phantom was imaged with a previously calibrated photostimulable phosphor system using standard bedside chest radiography protocols. The Bayesian technique was then used to process this image. BIE incorporates a radial exponential convolution scatter model with two adjustable parameters. In previous reports, these parameters were optimized to reduce the residual fraction of scattered radiation in the processed image. Here, the parameters were adjusted to evaluate the potential of BIE to reduce image noise. While the full width at half maximum of the scatter model was held constant, the magnitude was varied. Evaluation was based on residual scatter fractions and CNR. The magnitude of the kernel in the scatter model was varied from 0.0 to 2.5 in steps of 0.5. Previously, it was found that an ''ideal'' scatter kernel magnitude of 2.33 provided a minimum residual scatter fraction. This magnitude corresponds to the average scatter-to-primary ratio in the chest radiograph. As the magnitude was increased, the residual scatter fraction decreased and the CNR increased in both the lungs and the mediastinum. However, as the magnitude was decreased, the percent noise also decreased; therefore, a lower magnitude kernel reduces noise. By varying the magnitude of the kernel used, differing amounts of noise reduction and contrast enhancement can be obtained. These results demonstrate that Bayesian image estimation can be used to both increase contrast and decrease noise in digital chest radiography.
引用
收藏
页码:1255 / 1261
页数:7
相关论文
共 23 条
[1]   SCANNING MULTIPLE SLIT ASSEMBLY - PRACTICAL AND EFFICIENT DEVICE TO REDUCE SCATTER [J].
BARNES, GT ;
BREZOVICH, IA ;
WITTEN, DM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1977, 129 (03) :497-501
[2]   CHARACTERIZATION OF THE POINT SPREAD FUNCTION AND MODULATION TRANSFER-FUNCTION OF SCATTERED RADIATION USING A DIGITAL IMAGING-SYSTEM [J].
BOONE, JM ;
ARNOLD, BA ;
SEIBERT, JA .
MEDICAL PHYSICS, 1986, 13 (02) :254-256
[3]   STUDIES OF PERFORMANCE OF ANTISCATTER GRIDS IN DIGITAL RADIOGRAPHY - EFFECT ON SIGNAL-TO-NOISE RATIO [J].
CHAN, HP ;
LAM, KL ;
WU, YZ .
MEDICAL PHYSICS, 1990, 17 (04) :655-664
[4]   MAXIMUM LIKELIHOOD FROM INCOMPLETE DATA VIA EM ALGORITHM [J].
DEMPSTER, AP ;
LAIRD, NM ;
RUBIN, DB .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-METHODOLOGICAL, 1977, 39 (01) :1-38
[5]   SINGLE-EXPOSURE DUAL-ENERGY COMPUTED RADIOGRAPHY - IMPROVED DETECTION AND PROCESSING [J].
ERGUN, DL ;
MISTRETTA, CA ;
BROWN, DE ;
BYSTRIANYK, RT ;
SZE, WK ;
KELCZ, F ;
NAIDICH, DP .
RADIOLOGY, 1990, 174 (01) :243-249
[6]   BAYESIAN RESTORATION OF CHEST RADIOGRAPHS - SCATTER COMPENSATION WITH IMPROVED SIGNAL-TO-NOISE RATIO [J].
FLOYD, CE ;
BAYDUSH, AH ;
LO, JY ;
BOWSHER, JE ;
RAVIN, CE .
INVESTIGATIVE RADIOLOGY, 1994, 29 (10) :904-910
[7]   SCATTER COMPENSATION FOR DIGITAL CHEST RADIOGRAPHY USING MAXIMUM-LIKELIHOOD EXPECTATION MAXIMIZATION [J].
FLOYD, CE ;
BAYDUSH, AH ;
LO, JY ;
BOWSHER, JE ;
RAVIN, CE .
INVESTIGATIVE RADIOLOGY, 1993, 28 (05) :427-433
[8]   QUANTITATIVE RADIOGRAPHIC IMAGING USING A PHOTOSTIMULABLE PHOSPHOR SYSTEM [J].
FLOYD, CE ;
CHOTAS, HG ;
DOBBINS, JT ;
RAVIN, CE .
MEDICAL PHYSICS, 1990, 17 (03) :454-459
[9]   POSTERIOR BEAM-STOP METHOD FOR SCATTER FRACTION MEASUREMENT IN DIGITAL RADIOGRAPHY [J].
FLOYD, CE ;
BAKER, JA ;
LO, JY ;
RAVIN, CE .
INVESTIGATIVE RADIOLOGY, 1992, 27 (02) :119-123
[10]   SCATTER COMPENSATION IN DIGITAL CHEST RADIOGRAPHY USING FOURIER DECONVOLUTION [J].
FLOYD, CE ;
BEATTY, PT ;
RAVIN, CE .
INVESTIGATIVE RADIOLOGY, 1989, 24 (01) :30-33