MEASUREMENT OF CARDIAC LEFT-VENTRICULAR PRESSURE IN CONSCIOUS RATS USING A FLUID-FILLED CATHETER

被引:18
作者
SCHENK, J [1 ]
HEBDEN, A [1 ]
MCNEILL, JH [1 ]
机构
[1] UNIV BRITISH COLUMBIA,FAC PHARMACEUT SCI,VANCOUVER V6T 1Z3,BC,CANADA
关键词
LEFT VENTRICULAR PRESSURE; CONSCIOUS RATS; CATHETER NATURAL FREQUENCY;
D O I
10.1016/1056-8719(92)90037-2
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
A fluid-filled catheter consisting of 100 cm of PE50 polyethylene tubing welded to 7 cm of PE10 polyethylene tubing (PE50/PE10) was constructed for the purpose of measuring the rate of left ventricular pressure development (+ dP/dt) in conscious, freely moving rats. Prior to in vivo experiments, four PE50/PE10 catheters were randomly selected, and their natural frequencies and damping ratios were determined using a square wave impact. The mean (n = 4), natural frequency of these catheters was shown to be 35.0 +/- 5.5 Hz, and the mean damping ratio was 0.83 +/- 0. 10. Natural frequency plotted against increasing PE50 tubing length was shown to have a slope of - 0.44 Hz/cm with a correlation coefficient of 0.99. The effect of the 7-cm PE10 tubing segment on the catheter damping ratio was also demonstrated. One of the four PE50/PE10 type catheters exhibited a damping ratio of 0.74 +/- 0.09. When the 7-cm PE10 tube was removed, the damping ratio was reduced to 0.31 +/-0.04. Left ventricular + dP/dt obtained in conscious rats with a PE50/PE10 catheter (n = 7; 6300 +/- 300 mmHg/sec) was significantly less than the + dP/dt obtained using a 100-cm PE50 catheter (n = 6; 9400 +/- 400 mmHg/sec). The results of this study make it clear that the PE50/PE10 catheter is suitable for the measurement of left ventricular + dP/dt in the conscious rat, and that catheter design has a profound influence on both the catheter natural frequency and damping ratio.
引用
收藏
页码:171 / 175
页数:5
相关论文
共 11 条
[1]   INTEGRATIVE NATURE AND TIME COURSE OF CARDIOVASCULAR ALTERATIONS IN THE DIABETIC RAT [J].
DOWELL, RT ;
ATKINS, FL ;
LOVE, S .
JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1986, 8 (02) :406-413
[2]   MULTIPLE SIMULTANEOUS DETERMINATIONS OF HEMODYNAMICS AND FLOW DISTRIBUTION IN CONSCIOUS RAT [J].
FLAIM, SF ;
NELLIS, SH ;
TOGGART, EJ ;
DREXLER, H ;
KANDA, K ;
NEWMAN, ED .
JOURNAL OF PHARMACOLOGICAL METHODS, 1984, 11 (01) :1-39
[3]  
GARDINER SM, 1980, MED BIOL, V58, P232
[4]  
GEDDES LA, 1970, DIRECT INDIRECT MEAS
[5]  
Hansen A. T., 1949, ACTA PHYSIOL SCA S68, V19, P1
[6]  
HEBDEN RA, 1990, AM J PHYSIOL, V258, pR1042
[7]   RELATIONSHIP BETWEEN ATRIAL GRANULARITY AND RELEASE OF ATRIAL NATRIURETIC FACTOR IN RATS WITH DIABETES-MELLITUS [J].
HEBDEN, RA ;
TODD, ME ;
MCNEILL, JH .
AMERICAN JOURNAL OF PHYSIOLOGY, 1989, 257 (04) :R932-R938
[8]   VASCULAR INTRACARDIAC CATHETERIZATION TECHNIQUE FOR MULTIPHASIC EVALUATION OF RAT-HEART INVIVO [J].
KOPP, SJ .
TOXICOLOGY AND APPLIED PHARMACOLOGY, 1983, 70 (02) :273-282
[9]  
McDonald D. A., 1974, BLOOD FLOW ARTERIES
[10]  
PAULSON DJ, 1987, J PHARMACOL EXP THER, V240, P529