SPARK ABLATION AS SAMPLING DEVICE FOR INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRIC ANALYSIS OF LOW-ALLOYED STEELS

被引:12
作者
COEDO, AG
DORADO, MT
FERNANDEZ, B
机构
[1] Centro Nacional de Investigaciones Metalurgicas (CSIC), 28040 Madrid
关键词
SPARK ABLATION; INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; LOW-ALLOYED STEELS ANALYSIS;
D O I
10.1039/ja9951000859
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the direct elemental analysis of low-alloyed steels is described. Spark ablation coupled with excitation in a 1.5 kW argon inductively coupled plasma was applied to the direct determination of Mn, Si, P, Al, B, Co, Nb, V, Zr and Sb in low-alloyed steels by mass spectrometry, For aerosol production, a medium-voltage spark at a high repetition rate (500 V, 400 Hz) was used. A cyclone device was designed to control the aerosol particle size; with this system the total amount of material reaching the plasma is low (about 4 mu g min(-1)) and a more uniform particle size distribution is achieved, thereby improving the precision of the measurements and preventing deposition of material in the torch injector and sampling cone, Particles re-collected downstream of the cyclone device were mostly spherical in shape, with a particle diameter of less than 1.0 mu m. When using optimized operating parameters, and a low abundance iron isotope (Fe-58) as an internal standard, relative standard deviations at concentration levels of ten times the detection limit were below 2.5%, Detection limits (3s) for Nb, V, Zr, Mn, Co, Al and Sb ranged from 0.08 to 0.32 mu g g(-1); for Si, B and P, detection limits were 1.5, 2.1 and 2.5 mu g g(-1), respectively. These values are comparable to those obtained for continuous-flow sample aspiration of a 0.05% m/v Fe matrix solution, The technique was applied to the analysis of the Bureau of Analysed Samples certified reference materials SS-456/1 to SS-460/1, carbon steels (residual series).
引用
收藏
页码:859 / 863
页数:5
相关论文
共 13 条
[1]   DETERMINATION OF ALUMINUM, TITANIUM, NIOBIUM AND VANADIUM IN LOW-ALLOYED STEELS BY SPARK ABLATION COUPLED WITH INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
COEDO, AG ;
LOPEZ, MTD ;
SECO, JLJ ;
COBO, IG .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (01) :11-14
[2]  
GUSTAVSSON I, 1992, 3RD P INT C PROGR AN, P193
[3]   ANALYSIS OF CONDUCTING SOLIDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH SPARK ABLATION [J].
JAKUBOWSKI, N ;
FELDMANN, I ;
SACK, B ;
STUEWER, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (02) :121-125
[4]  
JARVIS KE, 1992, HDB INDUCTIVELY COUP, P280
[5]   ARC NEBULIZATION FOR ELEMENTAL ANALYSIS OF CONDUCTING SOLIDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
JIANG, SJ ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1986, 58 (08) :1739-1743
[6]  
KUSS HM, 1994, ATOM SPECTROSC, V15, P148
[7]  
KUSS HM, 1992, CHEM LISTY, V86, P623
[8]  
KUSS HM, 1992, 3RD P INT C PROGR AN, P201
[9]  
MOORE JAF, 1993, PLASMA SOURCE MASS S, V2, P150
[10]  
PAUL M, 1989, FRESEN Z ANAL CHEM, V334, P680